Z-0035-2012

CyberKnife Robotic Radiosurgery System, Synchrony Respiratory Tracking System (Standard Treatment Couch)...

Accuray Incorporated / initiated 2011-09-12 / Terminated / root cause: manufacturing process

Accuray Incorporated began a recall of CyberKnife Robotic Radiosurgery System, Synchrony Respiratory Tracking System (Standard Treatment Couch). Synchrony Respiratory Tracking System, a Radiation... on . The reason FDA records is "A report and investigation determined that the set screws, which were part of the retrofit of a previous recall were incorrectly engaged which allowed rotation and subsequent descent of the assembly". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0035-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Accuray Incorporated
Product
CyberKnife Robotic Radiosurgery System, Synchrony Respiratory Tracking System (Standard Treatment Couch). Synchrony Respiratory Tracking System, a Radiation Therapy Device, Model Numbers: 026429, 028185. Product is manufactured and distributed by Accuray, Inc., Sunnyvale, CA
Product code
IYE Accelerator, Linear, Medical
Reason for recall
A report and investigation determined that the set screws, which were part of the retrofit of a previous recall were incorrectly engaged which allowed rotation and subsequent descent of the assembly.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Urgent Device Correction Letters were sent to all sites affected by the retrofit on September 12, 2011 via Fed Ex. The letter identified the affected product and the description of the potential problem. Customers were asked to inspect their Synchrony System by following the instructions provided. If the lower extension feels loose, customers should contact Accuray Customer Support immediately to schedule service of the system. The letter also states that all affected systems will be inspected and corrected, if necessary by Accuray to prevent the potential problem. Accuray Customer Support will contact customers to schedule the inspection. Questions or concerns regarding the issue should be directed to Accuray Customer Support at 1-877-8667 for US customers and 1-408-716-4700 for customers outside of the US, or [email protected].
Quantity in commerce
236 total devices
Distribution
Worldwide Distribution
Product codes and lots
Model numbers 026429 and 028185. All units