"***REF BLU-LS9-5E***4.7 mm x 20 mm Fluted Drum, Side Cut Only. Use with B-BLUE-S Attachment...
The Anspach Effort, Inc. began a recall of "***REF BLU-LS9-5E***4.7 mm x 20 mm Fluted Drum, Side Cut Only. Use with B-BLUE-S Attachment. ***ANSPACH***Sterile, Sterilized using Irradiation***Rx Only... on . The reason FDA records is "The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0094-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- The Anspach Effort, Inc.
- Product
- "***REF BLU-LS9-5E***4.7 mm x 20 mm Fluted Drum, Side Cut Only. Use with B-BLUE-S Attachment. ***ANSPACH***Sterile, Sterilized using Irradiation***Rx Only ***Made in the USA***Manufacturer: The Anspach Effort, Inc. 4500 Riverside Drive, Palm Beach Gardens, FL 33410 USA Tel: (800) 327-6887 / + 1 561 627 1080***Anspach Europe Ltd. Aston Court, Kingsmead Business Park Frederick Place High Wycombe, HP11 1LA United Kingdom Tel: +441494 616126***" Lot number: 319461, B093004589. Product Usage: Cutting and shaping bone including bones of the spine and cranium.
- Product code
- HTT Burr, Orthopedic
- Reason for recall
- The Anspach Effort, Inc. initiated a recall on May 6, 2008 for their Sterile Bone Cutting Burrs. All affected customers were notified of the recall by letter dated 5/29/2008.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The firm, Anspach sent an "URGENT PRODUCT REMOVAL" letter on May 29, 2008, to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to do the following: 1) Screen their inventory immediately for all products listed on the attached "Product Replacement Form." Upon visual inspection if the pouch is sealed on both ends it is sterile. If one end of the pouch is not sealed it is non-sterile and should be returned for credit or replacement. 2) Complete the "Product Replacement Form" by indicating the quantity returned, and whether the customer would prefer to have their returned devices replaced, or their account credited. 3) Place products to be returned in a box and mark the box with the Return Authorization Number on the "Product Replacement Form." 4) Use the Anspach FedEx account number listed on the form to return the unused devices with the completed "Product Replacement Form." Following receipt of the form, Anspach will process the request immediately. For questions related to the product replacement, please contact Customer Service at 800-327-6887. For questions related to the Voluntary Removal, please contact Anspach Regulatory Affairs and Quality Assurance at 800-327-6887.
- Quantity in commerce
- 12
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of: CA, CO, FL, GA, IA, ID, SC, MA, MD, MI, MO, MN, MS, NY, OH, OR, PA, TN, TX, and WI, and the countries of: Africa, Netherlands, Panama, South Korea, Spain, and United Kingdom.
- Product codes and lots
- Lot number: 319461, B093004859.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
