Z-0003-2012

Cochlear Nucleus CI512 Cochlear Implant, REF Z209051, New with Nucleus 5, Sterile EO, Made in Australia...

Cochlear Americas / initiated 2011-09-16 / Terminated / root cause: design

Cochlear Americas Inc. began a recall of Cochlear Nucleus CI512 Cochlear Implant, REF Z209051, New with Nucleus 5, Sterile EO, Made in Australia, Cochlear Ltd 14 Mars Road, Lane Cove 2066, Australia... on . The reason FDA records is "Cochlear implant may shut down and cease to function". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0003-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cochlear Americas
Product
Cochlear Nucleus CI512 Cochlear Implant, REF Z209051, New with Nucleus 5, Sterile EO, Made in Australia, Cochlear Ltd 14 Mars Road, Lane Cove 2066, Australia. The cochlear implant is intended to restore a level of auditory sensation via the electrical stimulation of the auditory nerve.
Product code
MCM Implant, Cochlear
Reason for recall
Cochlear implant may shut down and cease to function.
Root cause tag
design
FDA root cause
Component design/selection
Action
Cochlear Americas, Inc. sent an "URGENT MEDICAL DEVICE RECALL" letter dated September 16, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter instructs customers to examine their inventory and quarantine the affected product. A recall response form was attached to the letter for customers to complete and return. Contact Cochlear Clinic Customer Service at 877-883-3101 for questions concerning this notice.
Quantity in commerce
33,645 units (global); 25,516 (implanted); 1,741 remaining in U.S. Market for implantation.
Distribution
Worldwide Distribution-USA (nationwide) and countries in Europe and Australia.
Product codes and lots
Part Number Z209051 all codes.