SKYLight Gamma Camera System; SKYLIGHT, 3/8",SPECT; Manufactured by Philips Medical Systems (Cleveland), Inc...
Philips Medical Systems began a recall of SKYLight Gamma Camera System; SKYLIGHT, 3/8",SPECT; Manufactured by Philips Medical Systems (Cleveland), Inc. 595 Miner Road, Cleveland, OH 44143. The SKYLight... on . The reason FDA records is "A mechanical malfunction of the button on the hand controller on the Philips Precedence and SKYLight SPECT systems may get stuck and cause unintended motion of a detector". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0249-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- SKYLight Gamma Camera System; SKYLIGHT, 3/8",SPECT; Manufactured by Philips Medical Systems (Cleveland), Inc. 595 Miner Road, Cleveland, OH 44143. The SKYLight Gamma Camera System is intended to product images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- A mechanical malfunction of the button on the hand controller on the Philips Precedence and SKYLight SPECT systems may get stuck and cause unintended motion of a detector.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Philips Medical Systems sent a "Customer Information" letter dated September 27, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter lists additional precautions the customers should use while operating the device. Philips will notify customers for replacing the affected product through an Action for Performance-Proactive Field Change Order to resolve the issue. Contact Philips Customer Care Solutions at 1-800-722-9377, option 5, for questions concerning this notice.
- Quantity in commerce
- 658 - all units in recall
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico.
- Product codes and lots
- Model numbers: 2160-3000A 4535-600-66661