Z-0150-2012

RSP Humeral Socket Reamer (Size 32 SML): DJO Surgical Instruments and Instrument Cases DJO surgical Encore...

Encore Medical, LP / initiated 2011-09-28 / Terminated / root cause: design

Encore Medical, Lp began a recall of RSP Humeral Socket Reamer (Size 32 SML): DJO Surgical Instruments and Instrument Cases DJO surgical Encore Medical, L.P. 9800 Metric Blvd, Austin, TX 78758... on . The reason FDA records is "DJO Surgical has determined that the alignment pins cannot be consistently cleaned using the technique provided in the cleaning/sterilization section of the instrument IFU". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0150-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
RSP Humeral Socket Reamer (Size 32 SML): DJO Surgical Instruments and Instrument Cases DJO surgical Encore Medical, L.P. 9800 Metric Blvd, Austin, TX 78758 Contact information: DJO Surgical, ATTN: Customer Service, 9800 Metric Blvd, Austin, TX 78758, 1-800-456-8696. The product is intended to be used to ream prepare the humeral bone for accepting the implant. It has a spring loaded alignment pin on the proximal (reaming) end used in both alignment of the reamer and as an indication that the reamer has been seated fully.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
DJO Surgical has determined that the alignment pins cannot be consistently cleaned using the technique provided in the cleaning/sterilization section of the instrument IFU.
Root cause tag
design
FDA root cause
Device Design
Action
DJO Surgical sent an URGENT FIELD SAFETY NOTICE letter dated September 28, 2011, to all affected customers. The letter identified the product. the problem, and the action to be taken by the customer. Customers were instructed to return the RSP Humeral Socket Reamers and RSP Baseplate Rim Planers. The RSP Humeral Socket Reamers will be reworked and returned to the customer. The Baseplast Rim Planer will be replaced when a suitable design is available. During the interim until instruments are returned, the use of FMP acetabular reamers size 42mm, 43mm, 45mm (DJO PIN 803-05-226, 227, 200) for the humeral preparation was validated to be effective to prepare the humerus for the implant. Customers were instructed to immediately notify the hospitals and surgeons using these devices and document it on the Field Safety Notice Response Form. For questions regarding this recall call 512-832-9500.
Quantity in commerce
399 Reamers
Distribution
Worldwide Distribution - USA including AR, AZ, CA, CO, FL,GA, ID, IN, IL, LA, MA, MD, ME, MI, MO, NJ, NM, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, and WI and Internationally toGermany, South America, Italy, Saudi Arabia, United Kingdom, and Canada.
Product codes and lots
All lots are being recalled.  DJO Surgical Part Number 804-02-013, 804-02-014, and 804-02-015