Z-0420-2012

Single Use PERPOS PLS System, Catalog #9045-01 Catalog#9045-01 for Single Use PERPOS PLS System- a sterile...

Interventional Spine, Inc. / initiated 2008-02-22 / Terminated / root cause: supplier

Interventional Spine Inc began a recall of Single Use PERPOS PLS System, Catalog #9045-01 Catalog#9045-01 for Single Use PERPOS PLS System- a sterile, medical device kit consisting of the components... on . The reason FDA records is "The recall was intiated because Interventional Spine has confirmed that the PERPOS 2-in-1 Drill (6070) manufactured by a contract manufacturer for Interventional Spine, Inc., may break during use". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-0420-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Interventional Spine, Inc.
Product
Single Use PERPOS PLS System, Catalog #9045-01 Catalog#9045-01 for Single Use PERPOS PLS System- a sterile, medical device kit consisting of the components: PERPOS Access Needle, PERPOS K-Wires, PERPOS 2-in-1 Drill, PERPOS Tap, PERPOS Device Driver, 4.5mm BONE-LOK PLS Implant, and PERPOS Pull Pin Remover. The kit is used by medical professions to stablize the spine as an aid to fusion through immobilization of the facet joints.
Product code
HTW Bit, Drill
Reason for recall
The recall was intiated because Interventional Spine has confirmed that the PERPOS 2-in-1 Drill (6070) manufactured by a contract manufacturer for Interventional Spine, Inc., may break during use.
Root cause tag
supplier
FDA root cause
Pending
Action
The recall communication was initiated on February 22, 2008 with Interventional Spine forwarding an Urgent Recall Notification to all their customers (via email). The notification informed the customers of the problem identified and instructions for the recall. If you have any questions, contact the Vice President, Quality, Regulatory and Clinical Affairs at (949) 525-1493 or email: [email protected].
Quantity in commerce
65
Distribution
Worldwide distribution: USA (nationwide) including states of: CA, FL, MO, NY, OH, PA, TN, WA, and WI; and country of: Europe.
Product codes and lots
Lot#090907A