Z-0154-2012

37% Etching Gel-1ml syringe 20 pk, part # T06A The intended use of these devices is to etch tooth...

Sybron Dental Specialties, Inc. / initiated 2011-03-31 / Terminated / root cause: manufacturing process

Kerr/pentron Dba Kerr Corporation And Pentron Clinical began a recall of 37% Etching Gel-1ml syringe 20 pk, part # T06A The intended use of these devices is to etch tooth dentin/enamel in order to prepare the tooth surface prior to... on . The reason FDA records is "Pentron Clinical is voluntarily recalling all lots of products labeled as 37% Phosphoric Acid Etching Gel produced since April 2007 because the product continues to extrude out of the syringe even after pressure is no...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0154-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Sybron Dental Specialties, Inc.
Product
37% Etching Gel-1ml syringe 20 pk, part # T06A The intended use of these devices is to etch tooth dentin/enamel in order to prepare the tooth surface prior to the application of a bonding agent.
Product code
EBF Material, Tooth Shade, Resin
Reason for recall
Pentron Clinical is voluntarily recalling all lots of products labeled as 37% Phosphoric Acid Etching Gel produced since April 2007 because the product continues to extrude out of the syringe even after pressure is no longer applied, and it contains a higher percentage of phosphoric acid than that indicated on the label.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, Pentron Clinical, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated May 31, 2011, with attached Return Form (via USPS 1st class mail) to all their consignees/customers. The letter describes the product, problem and action to be taken. All consignees/customers were instructed to complete the Recall Return Form via fax to: 1-877-677-8844 and to return any affected product in their inventory. Affected products will have expiration dates between April 2011 and April 2015. Consignees/customers with questions can contact Pentron Clinical Customer Service at (800) 551-0283 directly to handle the arrangements of a quick return and replacement.
Quantity in commerce
41,744 units
Distribution
Worldwide distribution: USA (nationwide) and countries of: Armenia, Austria, Australia, Antigua & Barbuda, Barbados, Belarus, Bulgaria, Canada, Czech Republic, Croatia, Cyprus, Ecuador, El Salvador, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kenya, Latvia, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Poland, Romania, Russia, St. Vincent & Grenadine, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, Ukraine, UK, and Vietnam.
Product codes and lots
all lots of products labeled as 37% Phosphoric Acid Etching Gel produced since April 2007