Elekta Synergy XVI 4.5
Elekta, Inc. began a recall of Elekta Synergy XVI 4.5 Product Usage: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner on . The reason FDA records is "If "Confirm" or "Unconfirm" settings are clicked during the "Terminated Checking" phase of treatment delivery, Desktop Pro cancels the loaded beam and sets the Beam MUs to zero. Desktop Pro then sends the delivered MUs...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0305-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Elekta Synergy XVI 4.5 Product Usage: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- If "Confirm" or "Unconfirm" settings are clicked during the "Terminated Checking" phase of treatment delivery, Desktop Pro cancels the loaded beam and sets the Beam MUs to zero. Desktop Pro then sends the delivered MUs to MOSAIQ a a zero value.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Elekta sent an IMPORTANT NOTICE letter dated March 9, 2011 to affected customers. The notice identified the product, the problem, and the corrective action to be taken by the customer. The notice advise all customers to follow any instructions or recommendations covered in the Notice. For questions call +44(0)1293 654200.
- Quantity in commerce
- 56
- Distribution
- Nationwide Distribution (USA) including the states of: AZ, CA, GA, IL, LA, MD, MA, MI, MN, NJ, NY, NC, OH, PA, TN, VA, WI and Puerto Rico
- Product codes and lots
- 152019, 152448, 151076, 151954, 152454, 152117, 152452, 152393, 152394, 151942, 152163, 151955, 152176, 152222, 151931, 152388, 151141, 152365, 152303, 152358, 152353, 152359, 152352, 152362, 152305, 152468, 152481, 152469, 151812, 152432, 151260, 151952, 152370, 151396, 152301, 152248, 152425, 151155, 151156, 151157, 151412, 105984, 152307, 152308. 151443, 151545, 151628, 152207, 152271, 152373, 152371, 151080, 152211, 152158, 152392, 152331