Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0283-2012

Siemens Medical Solutions USA, Inc., Symbia S Series SPECT System; Single-Photon Emission Computed...

Siemens Medical Solutions USA, Inc. / initiated 2011-10-14 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc. began a recall of Siemens Medical Solutions USA, Inc., Symbia S Series SPECT System; Single-Photon Emission Computed Tomography. Indications for use: SPECT + CT: Perform scans... on . The reason FDA records is "Errors could potentially impact image interpretation. The acquisition software is exhibiting a defect in the energy window position when an isotope is peaked. There is a potential patient safety issue when running syngo...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0283-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Medical Solutions USA, Inc., Symbia S Series SPECT System; Single-Photon Emission Computed Tomography. Indications for use: SPECT + CT: Perform scans and nuclear imaging studies with the same instrument. To obtain attenuation corrected images and to provide registration of anatomical and physiological images within the patient's anatomy.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
Errors could potentially impact image interpretation. The acquisition software is exhibiting a defect in the energy window position when an isotope is peaked. There is a potential patient safety issue when running syngo MI Applications version VA60A with Foresight detectors firmware version v1.5.9.12 on the Symbia systems.
Root cause tag
software
FDA root cause
Component change control
Action
Siemens Medical Solutions USA, Inc. sent an "Urgent Field Correction Recall" letter dated October 14, 2011 to all affected customers. The letters informed customers of the acquisition software defect in the energy window position when an isotope is peaked, which could potentially impact image interpretation, and that the AEB firmware version on their imaging system has been changed to a previously released version until acquisition software service pack is available. For questions or assistance they were instructed to contact Siemens Medical Solutions USA, Inc. at (800) 888-7436.
Quantity in commerce
2 units
Distribution
Florida, Illinois and Texas
Product codes and lots
Symbia S Series systems, serial numbers 1236 and 1555, both with syngo MI Applications version VA60A and Foresight Detector AEB firmware version v1.5.9.12