Z-0311-2012

Roche Cobas c 111 Analyzer The Roche Cobas c 111 analyzer is an in-vitro diagnostic analyzer capable of...

Roche Diagnostics GmbH / initiated 2011-10-18 / Terminated / root cause: software

Roche Diagnostics Operations, Inc. began a recall of Roche Cobas c 111 Analyzer The Roche Cobas c 111 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and... on . The reason FDA records is "Internal investigations have revealed a potential printer malfunction on the Cobas c 111 analyzer, where one or more numbers or characters can be missing on the hard copy result printout, or the printer may skip leaving...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0311-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche Cobas c 111 Analyzer The Roche Cobas c 111 analyzer is an in-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Internal investigations have revealed a potential printer malfunction on the Cobas c 111 analyzer, where one or more numbers or characters can be missing on the hard copy result printout, or the printer may skip leaving blanks which may result in an incorrect print layout. The missing number of characters can be in any position, therefore it could be in the middle of a result, (e.g. 48.53 mmol in
Root cause tag
software
FDA root cause
Software design
Action
Roche Diagnostics Operations, Inc sent a Urgent Medical Device Correction dated October 18, 2011, via UPS Ground (receipt required) to all affected customers and one distribution center, with instructions to disable the Auto Print Result feature until new software is available. All printed results will need to be confirmed against the results indicated on the display screen. Questions are to be directed to Roche Diagnostics Technical Support at 1-800 428-2336,
Quantity in commerce
135
Distribution
Worldwide Distribution - USA including West Virginia and Texas.
Product codes and lots
Serial Numbers 50865, 50626, 50798, 50731, 50654, 50541, 51023, 50946, 50989, 50634, 50506, 50251, 50887, 50948, 50453, 50688, 51111, 50997, 51122, 50244, 50790, 50475, 50597, 50740, 50656, 50793, 51086, 50999, 51146, 50635, 50744, 50586, 50632, 50846, 1167, 50693, 50787, 50670, 50672, 51128, 50941, 50562, 50788, 50547, 51003, 50624, 50677, 50678, 50814, 51015, 50687, 50680, 50894, 50560, 51004, 50247, 50538, 51118, 50926, 50847, 50841, 51127, 50741, 50764, 50633, 51106, 51116, 50201, 51124, 50612, 51088, 50576, 51016, 50689, 50681, 50424, 51126, 50959, 50878, 50784, 50661, 50761, 50452, 50450, 50290, 50947, 51327, 50630, 50985, 50763, 50319, 50984, 50250, 50549, 50925, 50844, 50660, 50845, 50683, 50735, 51056, 50463, 50775, 50840, 50734, 50655, 50843, 50758, 50852, 50539, 50651, 50653, 50657, 50964, 50563, 50742, 50738, 50727, 50934, 50561, 50803, 51000, 51107, 50686, 50815, 50649, 50470, 50759, 50644, 51053, 50739, 50595, 51087, 50575, and 50679.