Z-0319-2012

Stereotactic Circular Collimator

Elekta Limited / initiated 2011-03-25 / Terminated

Elekta, Inc. began a recall of Stereotactic Circular Collimator Product Usage: This device is intended to hold a patient's head in a fixed position and to localize and center the output of a... on . The reason FDA records is "Recent newspaper articles outlined improper use of SRS Cone Collimator accessories that injured patients on Brainlab and Varian systems". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0319-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Stereotactic Circular Collimator Product Usage: This device is intended to hold a patient's head in a fixed position and to localize and center the output of a linear accelerator (UNAC) to allow radiotherapy of brain tumors and other types of cerebral lesions.
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
Recent newspaper articles outlined improper use of SRS Cone Collimator accessories that injured patients on Brainlab and Varian systems.
FDA root cause
Under Investigation by firm
Action
Elekta sent an IMPORTANT NOTICE letter dated March 25, 2011 to all affected customers. The letter identified the affected product, the problem, key safety procedures and information to users for safe operation of the Stereotactic Circular Collimator. Elekta will issue a Mandatory Action Field Change Order which will include an updated Circular Collimator Instructions for Use - 102022201 and new labels. The notice instructs Users to file this document in the Important Notice section of the appropriate User Manual. Also, Elekta is offering some free WebEx sessions which will include the procedures and recommended operation of these types of accessories. Customers are instructed to register at www.elekta.com/SafeAccessories.
Quantity in commerce
89 devices
Distribution
Nationwide Distribution (USA) - including the states of: AZ, CA, FL, GA, MO, NJ, NM, NC, OR, TN, and WV
Product codes and lots
848, 849, 850, 851, 852, 853, 1032, 1033, 1034, 928, 929, 930, 931, 932, 801, 802, 803, 804, 805, 980, 981, 982, 983, 984, 889, 890, 891, 892, 893, 894, 908, 909, 910, 911, 912, 833, 834, 835, 836, 837, 843, 844, 845, 846, 847, 1017, 1018, 1019, 1020, 1021, 903, 904, 905, 906, 907, 883, 884, 885, 886, 933, 935, 936, 937, 940, 974, 975, 976, 977, 978, 941, 942, 943, 944, 945, 946, 947, 948, 949, 950, 1012, 1013, 1014, 1015, 1016, 1027, 1028, 1029, 1030, 1031