TC pCO2 Electrodes, Model/Part #945-655-E5260, 510(k) #K043003
Radiometer America Inc began a recall of TC pCO2 Electrodes, Model/Part #945-655-E5260, 510(k) #K043003. Product Usage: The device is intended for used with the firm's TCM series monitor. The... on . The reason FDA records is "Transcutaneous pCO2-electrodes were wrongly coded as: pO2 electrode on the hardware plug for the firm's TCM series monitors. This causes the monitor to interpret the electrode wrongly, causing it to fail all...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0341-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Radiometer America, Inc
- Product
- TC pCO2 Electrodes, Model/Part #945-655-E5260, 510(k) #K043003. Product Usage: The device is intended for used with the firm's TCM series monitor. The E260mpCO2 electrode is a non-invasive device, used with the TCM series monitor to measure transcutaneous pCO2 (through the skin).
- Product code
- LKD Monitor, Carbon-Dioxide, Cutaneous
- Reason for recall
- Transcutaneous pCO2-electrodes were wrongly coded as: pO2 electrode on the hardware plug for the firm's TCM series monitors. This causes the monitor to interpret the electrode wrongly, causing it to fail all calibrations and never get ready for use.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Radiometer contacted the affected U.S. customer and informed them of the recall via telephone on September 27, 2011. Radiometer requested the return of the affected product received from the firm. For questions call (440) 871-0463.
- Quantity in commerce
- 4 units were distributed.
- Distribution
- Worldwide Distribution - USA (nationwide) in the state of San Diego, CA.