Z-0211-2012
IMPAX Cardiovascular Results Management (RM) Software Versions RM 2.04.37.04 to RM 7.8 SU2 The IMPAX...
AGFA Healthcare Corp. / initiated 2011-06-06 / Terminated / root cause: software
AGFA Corp. began a recall of IMPAX Cardiovascular Results Management (RM) Software Versions RM 2.04.37.04 to RM 7.8 SU2 The IMPAX Cardiovascular suite is a cardiovascular information... on . The reason FDA records is "Improper merging of IMPAX Cardiovascular Results Management (RM) reports causes the signed report to become unavailable". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0211-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- AGFA Healthcare Corp.
- Product
- IMPAX Cardiovascular Results Management (RM) Software Versions RM 2.04.37.04 to RM 7.8 SU2 The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Improper merging of IMPAX Cardiovascular Results Management (RM) reports causes the signed report to become unavailable.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- AGFA Healthcare sent an "URGENT FIELD SAFETY NOTICE" letter dated May 27, 2011, via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to carry out a merge post finalization of an RM reports. If it is necessary to modify a finalized report, AGFA recommends that they make the modification through one of the following actions: Amend the report through the RM interface, or Call AGFA Service to change the status of the report to "preliminary" and then proceed with the finalization of the report. AGFA issued an Acknowledgment, via FAX- Back that the information was received and understood to affected customers. For any questions customers were instructed to contact their AGFA Healthcare Service at 877-777-2432.
- Quantity in commerce
- 323
- Distribution
- Worldwide Distribution - USA (nationwide) and Canada
- Product codes and lots
- Software Versions: RM2.04.37.04 to RM 7.8 SU2 (RM 2.04.37.04 is a version prior to CV 7.2 release).