Z-0272-2012

IMPAX CardioVascular (CV) Admin Tool The IMPAX Cardiovascular suite is a cardiovascular information system...

AGFA Healthcare Corp. / initiated 2011-06-06 / Terminated / root cause: software

AGFA Corp. began a recall of IMPAX CardioVascular (CV) Admin Tool The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and... on . The reason FDA records is "Potential for misattribution of data and for a report to become unavailable when the user requests for it to be moved to another study that already has a report associated with it". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0272-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
AGFA Healthcare Corp.
Product
IMPAX CardioVascular (CV) Admin Tool The IMPAX Cardiovascular suite is a cardiovascular information system providing image archiving and image display and modality study-specific structured reporting.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Potential for misattribution of data and for a report to become unavailable when the user requests for it to be moved to another study that already has a report associated with it.
Root cause tag
software
FDA root cause
Software design
Action
An 'URGENT FIELD SAFETY NOTICE" was sent via FED-EX by June 6, 2011 to all the sites using the affected product. The letter described the issue and mitigation. The letter recommends that users not move reports to studies where a report already exists. If a report is being moved because the existing report is irrelevant to the study, then the existing report should be deleted prior to moving the new report to that study. If the report is being moved because the user wants to combine the content of the reports, then the "merge" function should be utilized. The letter also states that a software enhancement is currently available. Acknowledgment via FAX-Back was requested from the sites. Questions regarding the issue should be directed to customers' local Agfa HealthCare Service at 877-777-2432. The firm has identified 185 additional sites and an "URGENT FIELD SAFETY NOTICE" was sent on October 12, 2012.
Quantity in commerce
323 units
Distribution
Worldwide Distribution -- USA, Virgin Islands and Canada
Product codes and lots
Software Versions 1.00.26, 1.00.27, CV7.4.SU3 and CV7.8.