Z-0134-2012

Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump; Baxter Healthcare Corporation, Deerfield, IL...

Baxter Healthcare Corporation / initiated 2011-08-26 / Terminated / root cause: manufacturing process

Baxter Healthcare Corp. began a recall of Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump; Baxter Healthcare Corporation, Deerfield, IL 60015; Product Code # 2M8064. The Flo-Gard Volumetric... on . The reason FDA records is "The Flo-Gard infusion Pump was released to the customer with OOS values for the Air Sensor Calibrations with the Air in Tubing test". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0134-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump; Baxter Healthcare Corporation, Deerfield, IL 60015; Product Code # 2M8064. The Flo-Gard Volumetric Infusion Pumps are an electromechanical pump used for intravenous infusion of fluids at user-selected rates.
Product code
FRN Pump, Infusion
Reason for recall
The Flo-Gard infusion Pump was released to the customer with OOS values for the Air Sensor Calibrations with the Air in Tubing test.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
A Baxter Healthcare technician initially notified the sole customer on August 26, 2011 of the defective product and requested that the product be quarantined. On September 14, 2011 a technician visited the site to correct the problem observed on the Flo-Gard Infusion Pump. All required tests were performed and the results were found acceptable.
Quantity in commerce
1 unit
Distribution
Distributed only in Puerto Rico.
Product codes and lots
product code 2M8064, serial number 8110057FB