Z-0191-2012
Spectra Optia Apheresis System, REF: 61000, Caridian BCT, Lakewood, CO 80215. A blood component separator...
Terumo BCT, Ltd / initiated 2011-09-21 / Terminated / root cause: software
Caridian BCT, Incorporated began a recall of Spectra Optia Apheresis System, REF: 61000, Caridian BCT, Lakewood, CO 80215. A blood component separator used for therapeutic plasma exchanges on . The reason FDA records is "Possible higher than expected platelet loss following therapeutic plasma exchange or plasmapheresis". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0191-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo BCT, Ltd
- Product
- Spectra Optia Apheresis System, REF: 61000, Caridian BCT, Lakewood, CO 80215. A blood component separator used for therapeutic plasma exchanges.
- Product code
- LKN Separator, Automated, Blood Cell And Plasma, Therapeutic
- Reason for recall
- Possible higher than expected platelet loss following therapeutic plasma exchange or plasmapheresis.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- CaridianBCT sent a "Safety Alert" letter dated September 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm will be sending a recall letter to consignees in November 2011, stating they are voluntarily recalling the software and will provide an upgrade once the software has been approved. If you have any further questions please contact your CaridianBCT Representative, the CaridianBCT Support Center at 1-877-3-FYU-BCT ( US Toll free 1-877-299-4228 ) or (303) 231-HELP ) ( 1 (303) 231-4357, or your local CaridianBCT Customer Service office.
- Quantity in commerce
- 852 units (US: 333; Foreign: 519)
- Distribution
- Worldwide Distribution -- USA ( nationwide ) and the countries of Australia, Belgium, Canada, Chile, China, Colombia, Hong Kong, India, Japan, Malaysia, Mexico, Pakistan, Republic of Korea, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand and Vietnam.
- Product codes and lots
- Serial Numbers: 1P00100 - 1P00XXX