Z-0381-2012
UltraLite Headlight module-grey Model: 001328
Integra LifeSciences Corporation / initiated 2011-10-10 / Terminated / root cause: design
Integra LifeSciences Corp. began a recall of UltraLite Headlight module-grey Model: 001328 Product Usage: The fiber optic headlight system Is designed to deliver illumination from a high intensity light... on . The reason FDA records is "Headlight assembly (module) may become loose and separate from the headband assembly". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0381-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Integra LifeSciences Corporation
- Product
- UltraLite Headlight module-grey Model: 001328 Product Usage: The fiber optic headlight system Is designed to deliver illumination from a high intensity light source for surgical site illumination.
- Product code
- FST Light, Surgical, Fiberoptic
- Reason for recall
- Headlight assembly (module) may become loose and separate from the headband assembly
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Integra Life Sciences notified consignees by URGENT DEVICE RECALL letter dated October 10, 2011, via FedEx, email or telephone conference calls. The letter identified the affected product, the problem, and the actions to be taken. Customers were instructed to identify and return any affected product. Customers were requested to complete and return the attached Field Corrective Action Acknowledgement Form to Integra Surgical indicating receipt and review of this notification. For assistance and additional information contact your local and Integra Sales Representative or call Integra Surgical Customer Service at 1-800-431-1123.
- Quantity in commerce
- 60 units
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico, and the countries of: Africa, Algeria, Australia, Austria, Brazil, Canada, Check Republic, China, Chile, Costa Rica, Denmark, Dominican Republic, Ecuador, Ethiopia, Finland, France, Georgia, Germany, Greece Hungary, Guatemala, India, Ireland, Israel, Italy, Iraq, Japan, Jordan, Kenya, Kuwait, Lebanon, Lithuania, Macedonia, Mexico, Netherlands, New Zealand! Norway, Oman, Pakistan, Peru Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad, Turkey, Qatar, United Kingdom, Yemeni United Arab Emirates, Uruguay, and Venezuela.
- Product codes and lots
- All serial numbers