Z-0399-2012

VARIANT II Link Hemoglobin A1c Program Reorder Pack, 1600 tests, IVD, For the determination of Hemoglobin A1c...

Bio-Rad Laboratories Inc. / initiated 2011-10-25 / Terminated / root cause: manufacturing process

Bio-Rad Laboratories, Inc. began a recall of VARIANT II Link Hemoglobin A1c Program Reorder Pack, 1600 tests, IVD, For the determination of Hemoglobin A1c in human whole blood. Contents: Elution Buffer A... on . The reason FDA records is "It was discovered that the Variant II Turbo Link Calibrator Values card contains incorrect values for IFCC units. The numbers are missing a decimal point". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0399-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Rad Laboratories Inc.
Product
VARIANT II Link Hemoglobin A1c Program Reorder Pack, 1600 tests, IVD, For the determination of Hemoglobin A1c in human whole blood. Contents: Elution Buffer A , 1 x 2500 mL Elution Buffer B, 1 x 2100 mL Wash/Diluent Solution, 1 x 2500 mL Cartridge Set CD ROM - 1 Calibrator/Diluent Set Whole Blood Primer, 6 x 1 mL Sample vials, 2 x 100 Instruction Manual Bio-Rad Laboratories, Inc., 400 Alfred nobel Drive, Hercules, CA 94547-1803. Intended for the percent determination of hemoglobin A1c in human whole blood using ion-exchange high-performance liquid chromatography (HLCP).
Product code
KRZ Calibrator For Hemoglobin And Hematocrit Measurement
Reason for recall
It was discovered that the Variant II Turbo Link Calibrator Values card contains incorrect values for IFCC units. The numbers are missing a decimal point.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Bio-Rad Laboratories, Inc. sent a Medical Device Correction Notification letter dated October 25, 2011, via Federal Express to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The correction was initiated by the firm, customers were instructed to contact their regional Bio-Rad office for any questions. For questions regarding this recall call 510-741-6114.
Quantity in commerce
63
Distribution
Nationwide Distribution including MI, SC, OH, IL, MN, MD, VA, NY and WI.
Product codes and lots
Catalog number: 270-2716: Lot 70211322; 70211432, 70211440; exp 3/31/2012