Varian brand GammaMed 12i Accessory: Clamping adapters for bronchial catheters, 1.8mm diameter, GammaMed 12i...

Varian Medical Systems, Inc. / initiated 2011-10-06 / Terminated / root cause: design

Varian Medical Systems, Inc. Oncology Systems began a recall of Varian brand GammaMed 12i Accessory: Clamping adapters for bronchial catheters, 1.8mm diameter, GammaMed 12i; Ref: BT-00264, Model 2-H63; Product is... on . The reason FDA records is "The Product Models 1 & 2 have a design flaw problem where the clamping adapter can obstruct movement of the source cable in the event the bronchial catheter is dislodged from the connector". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0330-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
Varian brand GammaMed 12i Accessory: Clamping adapters for bronchial catheters, 1.8mm diameter, GammaMed 12i; Ref: BT-00264, Model 2-H63; Product is distributed by Varian Medical Systems Inc., Palo Alto, CA and Manufactured by Varian Medical Systems Haan GmbH, Bergische Strasse 16, D-42781 Haan, Germany. The Varian Medical System GammaMed Afterloader system is a remotely controlled intelligent robotic device intended to be used to deliver high dose-rate radiation therapy, in close proximity to and from within diseased tissues. Palliative care and treatments with curative intent to both benign and non-benign disease using interstitial, intracavitary, intraluminal, intra and inter-operative brachytherapy or combinations of any these treatment types are possible using this device.
Product code
JAQ System, Applicator, Radionuclide, Remote-Controlled
Reason for recall
The Product Models 1 & 2 have a design flaw problem where the clamping adapter can obstruct movement of the source cable in the event the bronchial catheter is dislodged from the connector.
Root cause tag
design
FDA root cause
Device Design
Action
Varian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION LETTER/FIELD SAFETY NOTICE" letter dated October 6, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Contact the Varian Help Desk at 1-888-827-4265 for questions regarding this notice.
Distribution
Worldwide Distribution--USA (nationwide).
Product codes and lots
CODES:   H630164 ,H630703 ,H630727 ,H630756 ,H630787 ,H630810 ,H630829 ,H630850 ,H630868 ,H630888 ,H630165 ,H630705 ,H630728 ,H630757 ,H630789 ,H630811 ,H630832 ,H630851 ,H630869 ,H630889 ,H630168 ,H630706 ,H630729 ,H630758 ,H630790 ,H630812 ,H630833 ,H630852 ,H630870 ,H630891 ,H630170 ,H630711 ,H630732 ,H630760 ,H630792 ,H630813 ,H630834 ,H630854 ,H630871 ,H630892 ,H630205 ,H630712 ,H630733 ,H630764 ,H630793 ,H630814 ,H630835 ,H630855 ,H630872 ,H630895 ,H630206 ,H630713 ,H630736 ,H630769 ,H630794 ,H630815 ,H630837 ,H630857 ,H630877 ,H630896 ,H630210 ,H630714 ,H630737 ,H630771 ,H630795 ,H630817 ,H630839 ,H630858 ,H630878 ,H630897 ,H630211 ,H630716 ,H630739 ,H630774 ,H630798 ,H630818 ,H630841 ,H630859 ,H630879 ,H630898 ,H630214 ,H630718 ,H630741 ,H630776 ,H630801 ,H630820 ,H630842 ,H630860 ,H630880 ,H630901 ,H630215 ,H630719 ,H630744 ,H630777 ,H630802 ,H630822 ,H630843 ,H630863 ,H630881 ,H630902 ,H630216 ,H630720 ,H630747 ,H630778 ,H630804 ,H630825 ,H630845 ,H630864 ,H630884 ,H630903 ,H630219 ,H630721 ,H630749 ,H630781 ,H630807 ,H630826 ,H630847 ,H630866 ,H630885 ,H630904 ,H630701 ,H630722 ,H630753 ,H630784 ,H630809 ,H630828 ,H630848 ,H630867 ,H630886