Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0297-2012

Multigent Vancomycin Calibrators; Manufactured by Microgenics Corp, Fremont, CA. Distributed by: Abbott...

Microgenics Corporation / initiated 2011-06-08 / Terminated / root cause: manufacturing process

Microgenics Corp began a recall of Multigent Vancomycin Calibrators; Manufactured by Microgenics Corp, Fremont, CA. Distributed by: Abbott Diagnostics, Abbott Park, IL Product Usage: The... on . The reason FDA records is "Complaint from the sole distributor that the third party control material is recovering high, and outside the published range". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0297-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microgenics Corporation
Product
Multigent Vancomycin Calibrators; Manufactured by Microgenics Corp, Fremont, CA. Distributed by: Abbott Diagnostics, Abbott Park, IL Product Usage: The Multigent Vancomycin calibrators are intended for use in the calibration of the multiagent vancomycin Vancomycin is a commonly used antibiotic to treat staphylococcus and other gram-positive bacterial infections.
Product code
DLJ Calibrators, Drug Specific
Reason for recall
Complaint from the sole distributor that the third party control material is recovering high, and outside the published range.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Microgenics sent a Product Removal notification letter dated June 15, 2011 to all affected customers. The letter identified the affected product, the problem, and the necessary actions to be taken. Customers were instructed to discontinue use and dispose of the affected product per local waste ordinances, review all patient results reported with the use of this lot, and retain a copy of this notification for their records. The letter states that if the affected product has been forwarded to another laboratory, a copy of the letter is to be provided to them. An alternative calibrator is available to customers impacted. It can be ordered by calling Thermo Fischer Scientific Customer Service at 1-800-232-3342. Questions regarding the information should be forwarded to Abbott Customer Service at 1-877-4ABBOTT.
Quantity in commerce
809 kits lot 59446762; 651 kits lot 59391741
Distribution
Nationwide Distribution (USA) in the state of ILLINOIS
Product codes and lots
Catalog number: 6E44-01; Lot 59446762, Exp 3/31/12; lot 59391741, Exp 12/31/2011.