Synthes Oscillating Saw Attachment/Large w/Key for Small Battery Drive. This surgical instrument motor and...
Synthes USA (HQ), Inc. began a recall of Synthes Oscillating Saw Attachment/Large w/Key for Small Battery Drive. This surgical instrument motor and accessories are AC-powered, battery-powered or... on . The reason FDA records is "The potential exists for the Large Oscillating Saw Attachment to disengage from the Small Battery Drive handpiece. If the attachment disengages from the handpiece, there is a potential for injury to the patient and/or...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0348-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes Oscillating Saw Attachment/Large w/Key for Small Battery Drive. This surgical instrument motor and accessories are AC-powered, battery-powered or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. Accessories or attachments may include a bur, chisel (osteotome), dermabrasion brush, dermatome, drill bit, hammerhead, pin driver and saw blade.
- Product code
- HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment
- Reason for recall
- The potential exists for the Large Oscillating Saw Attachment to disengage from the Small Battery Drive handpiece. If the attachment disengages from the handpiece, there is a potential for injury to the patient and/or user.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Synthes sent a "NOTICE: MEDICAL DEVICE RECALL" letter dated October 17, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The firm requests that customers return their affected device(s) for servicing. Instructions are provided for customers to follow. Customers having questions regarding this notice should contact the firm at 610-719-5450.
- Quantity in commerce
- 5
- Distribution
- Nationwide Distribution including the states of CO, LA, NY, and OK.
- Product codes and lots
- Catalog number: 532.026 - All lots