Z-0348-2012

Synthes Oscillating Saw Attachment/Large w/Key for Small Battery Drive. This surgical instrument motor and...

Synthes GmbH / initiated 2011-10-17 / Terminated / root cause: design

Synthes USA (HQ), Inc. began a recall of Synthes Oscillating Saw Attachment/Large w/Key for Small Battery Drive. This surgical instrument motor and accessories are AC-powered, battery-powered or... on . The reason FDA records is "The potential exists for the Large Oscillating Saw Attachment to disengage from the Small Battery Drive handpiece. If the attachment disengages from the handpiece, there is a potential for injury to the patient and/or...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0348-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes Oscillating Saw Attachment/Large w/Key for Small Battery Drive. This surgical instrument motor and accessories are AC-powered, battery-powered or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. Accessories or attachments may include a bur, chisel (osteotome), dermabrasion brush, dermatome, drill bit, hammerhead, pin driver and saw blade.
Product code
HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment
Reason for recall
The potential exists for the Large Oscillating Saw Attachment to disengage from the Small Battery Drive handpiece. If the attachment disengages from the handpiece, there is a potential for injury to the patient and/or user.
Root cause tag
design
FDA root cause
Device Design
Action
Synthes sent a "NOTICE: MEDICAL DEVICE RECALL" letter dated October 17, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The firm requests that customers return their affected device(s) for servicing. Instructions are provided for customers to follow. Customers having questions regarding this notice should contact the firm at 610-719-5450.
Quantity in commerce
5
Distribution
Nationwide Distribution including the states of CO, LA, NY, and OK.
Product codes and lots
Catalog number: 532.026 - All lots