Abbott brand IntraLase Femtosecond Laser System, Model 3; Catalog Numbers: 20004D/J/K, R20004J/K; Product is...
AMO Manufacturing USA, LLC began a recall of Abbott brand IntraLase Femtosecond Laser System, Model 3; Catalog Numbers: 20004D/J/K, R20004J/K; Product is manufactured and distributed by AMO Manufacturing... on . The reason FDA records is "The firm has identified a discrepancy in cut depth that may affect deep corneal incisions (e.g. 300 m or deeper) created using the Laser System". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0166-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Advanced Medical Optics, Inc.
- Product
- Abbott brand IntraLase Femtosecond Laser System, Model 3; Catalog Numbers: 20004D/J/K, R20004J/K; Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA Product Usage: Intended Use for: IntraLase Femtosecond (Model 2 and 3) and iFS Advanced Femtosecond Laser Systems. FS Laser Systems are ophthalmic surgical lasers used in certain patient populations: (1) In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea; (2) In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments; (3) In lamellar IEK [lntraLase Enabled Keratoplasty] and corneal harvesting ; (4) In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea; (5) In the creation of a lamellar cut/resection of the cornea for lamellar IEK for the creation of a penetrating cut/incision for penetrating IEK.
- Product code
- HNO Keratome, Ac-Powered
- Reason for recall
- The firm has identified a discrepancy in cut depth that may affect deep corneal incisions (e.g. 300 m or deeper) created using the Laser System.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Abbott Medical Optics Inc. (AMO) issued an "ADVISORY NOTICE" letter dated October 12, 2011 to all affected customers. The letter identifies the affected products, the problem, and the actions needed to be taken. Customers were instructed to exercise caution when making cuts that approach the corneal endothelium. The letter includes a 1-page Operator's Manual addendum that clarifies the recommendations. The letter states that AMO will conduct a field correction to adjust the EOF on the affected FS Laser Systems. The letter includes a response form to be completed and returned as per the instructions. Customers were instructed to notify all appropriate staff/personnel of the Advisory Notice. For any further questions regarding the information or recommendations, contact an AMO representative at 1-800-266-3375
- Quantity in commerce
- 1226 total laser systems
- Distribution
- Worldwide Distribution - USA (nationwide)
- Product codes and lots
- All units of these models (models 2, 3 and iFS) except model 1.