Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0167-2012

Abbott brand Advanced Femtosecond Laser Systems, Model iFS, Catalog Number: J20007D/J/K; Product is...

Advanced Medical Optics, Inc. / initiated 2011-10-12 / Terminated / root cause: manufacturing process

AMO Manufacturing USA, LLC began a recall of Abbott brand Advanced Femtosecond Laser Systems, Model iFS, Catalog Number: J20007D/J/K; Product is manufactured and distributed by AMO Manufacturing USA, LLC... on . The reason FDA records is "The firm has identified a discrepancy in cut depth that may affect deep corneal incisions (e.g. 300 m or deeper) created using the Laser System". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0167-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Medical Optics, Inc.
Product
Abbott brand Advanced Femtosecond Laser Systems, Model iFS, Catalog Number: J20007D/J/K; Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA Product Usage: Intended Use for: IntraLase Femtosecond (Model 2 and 3) and iFS Advanced Femtosecond Laser Systems. FS Laser Systems are ophthalmic surgical lasers used in certain patient populations: (1) In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea; (2) In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments; (3) In lamellar IEK [lntraLase Enabled Keratoplasty] and corneal harvesting ; (4) In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea; (5) In the creation of a lamellar cut/resection of the cornea for lamellar IEK for the creation of a penetrating cut/incision for penetrating IEK.
Product code
HNO Keratome, Ac-Powered
Reason for recall
The firm has identified a discrepancy in cut depth that may affect deep corneal incisions (e.g. 300 m or deeper) created using the Laser System.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Abbott Medical Optics Inc. (AMO) issued an "ADVISORY NOTICE" letter dated October 12, 2011 to all affected customers. The letter identifies the affected products, the problem, and the actions needed to be taken. Customers were instructed to exercise caution when making cuts that approach the corneal endothelium. The letter includes a 1-page Operator's Manual addendum that clarifies the recommendations. The letter states that AMO will conduct a field correction to adjust the EOF on the affected FS Laser Systems. The letter includes a response form to be completed and returned as per the instructions. Customers were instructed to notify all appropriate staff/personnel of the Advisory Notice. For any further questions regarding the information or recommendations, contact an AMO representative at 1-800-266-3375
Quantity in commerce
1226 total laser systems
Distribution
Worldwide Distribution - USA (nationwide)
Product codes and lots
All units of these models (models 2, 3 and iFS) except model 1.