Z-0349-2012

Foundation Tibial Broach P/N 801-01-013 Foundation Tibial Broach to be used in conjunction with the Tibial...

Encore Medical, LP / initiated 2011-10-26 / Terminated / root cause: manufacturing process

Encore Medical, Lp began a recall of Foundation Tibial Broach P/N 801-01-013 Foundation Tibial Broach to be used in conjunction with the Tibial Sizing Template and Tibial Broach Guide to prepare... on . The reason FDA records is "The strike plate may dislodge from the broach stem handle during impaction of the device". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0349-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
Foundation Tibial Broach P/N 801-01-013 Foundation Tibial Broach to be used in conjunction with the Tibial Sizing Template and Tibial Broach Guide to prepare the resected tibial plateau for acceptance of a keeled tibial Baseplate stem. P/N 801-01-013
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The strike plate may dislodge from the broach stem handle during impaction of the device.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
DJO Global sent an URGENT FIELD SAFETY NOTICE letter dated October 26, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to pass the information to all those who need to be aware within their organization or to any organization where the affected devices have been transferred. Customers should contact Customer Service to place a replacement order and receive an RMA number and return all affected devices using the RMA number. For any questions call (512) 834-6255.
Quantity in commerce
30 units
Distribution
Worldwide Distribution - USA including AR, AZ, CA, LA, MD, NY, TX, UT and Puerto Rico and the countries of United Kingdom and Italy
Product codes and lots
Lot#'s 35149L02, 36204L06, 37990L06, 38272L13, and 50479L12