Z-0310-2012

cobas c 311 Analyzer, Part Number: 04826876001 In-vitro diagnostic analyzer capable of performing clinical...

Roche Diagnostics GmbH / initiated 2011-10-26 / Terminated / root cause: software

Roche Diagnostics Operations, Inc. began a recall of cobas c 311 Analyzer, Part Number: 04826876001 In-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests... on . The reason FDA records is "The cobas c 311 analyzer software may incorrectly calculate the volume of the ISE Reference Electrolyte solution remaining on the analyzer. In some cases the incorrect results obtained will not be flagged with any data...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0310-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
cobas c 311 Analyzer, Part Number: 04826876001 In-vitro diagnostic analyzer capable of performing clinical chemistry, specific protein and electrolyte tests. Analytes are measured photometrically or turbidimetrically; the analyzer also has an optional ISE module for measuring sodium, potassium and chloride.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The cobas c 311 analyzer software may incorrectly calculate the volume of the ISE Reference Electrolyte solution remaining on the analyzer. In some cases the incorrect results obtained will not be flagged with any data alarms.
Root cause tag
software
FDA root cause
Software design
Action
October 26, 2011 Urgent Medical Device Corrections were sent to all direct accounts in the USA that have installed the cobas c 311 analyzer and included a faxback form to be completed and faxed to 1-877-766-7452. Consignees were instructed to temporarily replace the ISE Reference Electrolyte bottle based upon the frequency calculated from the maximum number of ISE samples run per day and activate the yellow threshold alarm on the system. Customers with question and concerns are directed to call Roche Diagnostics Technical Support at 1-800-428-2366, available 24-hours, 7 days a week.
Quantity in commerce
940 worldwide
Distribution
Worldwide Distribution -- USA including the states of Maryland and Texas.
Product codes and lots
Serial Numbers:1050-16, 1040-08, 1041-06, 0927-16, 0918-05, 1048-02, 0812-12, 1058-09, 0925-20, 1044-19, 1041-05, 1043-15, 1056-18, 0812-09, 1038-07, 0814-10, 1041-19, 1170-04, 0929-08, 0919-16, 0814-03, 0926-11, 1035-01, 1034-19, 1035-10, 1174-19, 1176-03, 1067-02, 0919-17, 0920-05, 0927-13, 1060-17, 1060-05, 0927-17, 1044-18, 0812-10, 0811-20, 1041-20, 1174-04, 0925-05, 1060-01, 1062-09, 1066-20, 1174-07, 0814-15, 0814-17, 1053-16, 1054-02, 1034-02, 1052-16, 1052-19, 0918-06, 1052-13, 1052-20, 1052-15, 1034-07, 1043-10, 0811-15, 1054-03, 1061-08, 1043-06, 0926-10, 1044-06, 1034-03, 1052-09, 1043-12, 1055-01, 0929-09, 1038-05, 1034-01, 0810-19, 1054-20, 1062-07, 1059-20, 0814-16, 1043-14, 1062-08, 1054-01, 0929-07, 1038-06, 1056-19, 1057-03, 1169-20, 0926-19, 1067-01, 1038-04, 0925-06, 1043-08, 1043-13, 0812-11, 1057-05, 1060-06, 0926-16, and 1034-06.