Z-0265-2012
Stryker Endoscopy 3.55 MM Super 90 S-SERFAS Energy Probe; electrosurgical device. Model number: 279-351-300
Stryker Corp. / initiated 2011-10-24 / Terminated / root cause: design
Stryker Endoscopy began a recall of Stryker Endoscopy 3.55 MM Super 90 S-SERFAS Energy Probe; electrosurgical device. Model number: 279-351-300 on . The reason FDA records is "An increased number of complaints for tip breaks were observed for the 3.5mm Super 90-S SERFAS Energy Suction Probes manufactured from June to August 2011". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0265-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker Endoscopy 3.55 MM Super 90 S-SERFAS Energy Probe; electrosurgical device. Model number: 279-351-300;
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- An increased number of complaints for tip breaks were observed for the 3.5mm Super 90-S SERFAS Energy Suction Probes manufactured from June to August 2011.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Stryker Endoscopy Urgent Medical Device Removal notifications datyed October 25, 2011, were sent to all consignees via traceable mail. Sales representatives and regional managers were notified by e-mail. Product is to be returned to Stryker Endoscopy using the pre-paid shipping label. Replacement product may be obtained by calling Stryker at 1-800-624-4422. The enclosed Acknowledgement receipt form is to be completed and returned. Questions or comments may be directed to Stryker at 1-800-624-4422 or e-mail at [email protected].
- Quantity in commerce
- 6307
- Distribution
- Worldwide distribution including USA, Australia, Canada, China, Greece, Hong Kong, India, Italy, Japan, France, Netherlands, Poland, South Africa, Spain, Sweden, Switzerland and the United Kingdom
- Product codes and lots
- Model number: 279-351-300; Lots 11161AE2 to 11241AE2, Lots are numbered sequentially using the middle three numbers, i.e 161-241.