Z-0279-2012

STERIS 5085 and 5085SRT Surgical Tables, Catalog Numbers: 5085- ST01-410-1; 5085SRT- ST0-1420-6

STERIS Corporation / initiated 2011-10-17 / Terminated / root cause: manufacturing process

Steris Corporation began a recall of STERIS 5085 and 5085SRT Surgical Tables, Catalog Numbers: 5085- ST01-410-1; 5085SRT- ST0-1420-6 on . The reason FDA records is "Some hydraulic column cylinders installed in certain 5085 and 5085SRT Surgical Tables were assembled by their supplier with incorrect snap rings and an anomaly was present in the tilt cylinder. This error may affect the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0279-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
STERIS 5085 and 5085SRT Surgical Tables, Catalog Numbers: 5085- ST01-410-1; 5085SRT- ST0-1420-6.
Product code
FQO Table, Operating-Room, Ac-Powered
Reason for recall
Some hydraulic column cylinders installed in certain 5085 and 5085SRT Surgical Tables were assembled by their supplier with incorrect snap rings and an anomaly was present in the tilt cylinder. This error may affect the user's ability to move the table top out of the full right tilt position.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On October 17, 2011, Steris sent 'Urgent Voluntary Field Correction Notices' to their customers via Fedex with tracking numbers for delivery confirmation. The letter identifies the affected product and notifies the user/customers of the problem with the surgical tables. The notification letter goes on to inform the customers that a STERIS Service Representative will be contacting them to arrange for the field correction of the affected unit(s). This involves the replacement of the hydraulic column cylinder of each surgical table(s).
Quantity in commerce
167 tables were distributed - 5085=108 and 5085SRT=59
Distribution
Nationwide distribution.
Product codes and lots
Tables identified with the following Codes are subject to correction: 0401510077 to 0421411110. (Note: not all tables manufactured in this range are affected by this voluntary field correction).