Z-0372-2012

Cincinnati Sub Zero, Steri-Probe Temperature Probe, Esophageal/Rectal Temperature Probe, Part Number: 39622...

Gentherm Medical, LLC / initiated 2011-09-26 / Terminated / root cause: manufacturing process

Cincinnati Sub-Zero Products Inc began a recall of Cincinnati Sub Zero, Steri-Probe Temperature Probe, Esophageal/Rectal Temperature Probe, Part Number: 39622, Catalog #491B. TA disposable temperature probe... on . The reason FDA records is "Reports that the Rectal/ Esophageal Temperature Probes are not reading correctly. This situation causes the Blanketrol device, which the probes connect to, to shut down operation and display a 'check probe' error". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0372-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gentherm Medical, LLC
Product
Cincinnati Sub Zero, Steri-Probe Temperature Probe, Esophageal/Rectal Temperature Probe, Part Number: 39622, Catalog #491B. TA disposable temperature probe used in conjunction with a measuring/monitoring device to continuously measure/monitor a patient's core body temperature.
Product code
BZT Stethoscope, Esophageal, With Electrical Conductors
Reason for recall
Reports that the Rectal/ Esophageal Temperature Probes are not reading correctly. This situation causes the Blanketrol device, which the probes connect to, to shut down operation and display a 'check probe' error.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Cincinnati Sub-Zero Products, Inc., issued an "Urgent Medical Device Field Action" letter on November 1, 2011 to all affected customers. The letter includes; description of product and problem, and a list of the affected lot numbers. The letter instructs the customers to discontinue use of the device. All unused product is to be returned to the Biomedical/Clinical Engineering and/or Service Department for an replacement and all used probes are to be disposed of in accordance with hospital policy. The customers are also instructed to complete and return an attached Recall Response Form ASAP. For additional information, customers may contact the Beatrice Washington at 1-800-989-7373 or (513) 772-8810.
Quantity in commerce
369,680
Distribution
Nationwide distribution and Canada.
Product codes and lots
The following lot codes are subject to recall: 12998, 12999, 19257, 19257A, 19644, 21536, 21537, 27446, 29112, 31081, 31082, 32209, 37281, 38522,  8996, D414723, D416607, D420120, D421702, D423801, D423908, D525212, D515408, D516607, D517108, D520709, D524107, D525502, D527632, D530518, D624011, D624012, D634834 D706611, D709119, D720450, D721118, D722015, and D727416.