Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0353-2012

Maquet Servo-S ventilator system Device Part number 66 40 440 Intended for treatment of and monitoring of...

Maquet Cardiovascular Us Sales, LLC / initiated 2011-09-19 / Terminated / root cause: supplier

Maquet Inc. began a recall of Maquet Servo-S ventilator system Device Part number 66 40 440 Intended for treatment of and monitoring of patients in range of neonates, infants and adults... on . The reason FDA records is "Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical...". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-0353-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Maquet Servo-S ventilator system Device Part number 66 40 440 Intended for treatment of and monitoring of patients in range of neonates, infants and adults with respiratory failure of respiratory insufficiency; used only by healthcare providers in hospitals or healthcare facilities and for in hospital transport.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical error message 24 (TE 24).
Root cause tag
supplier
FDA root cause
Vendor change control
Action
The firm, Maquet Getinge Group, sent an " PRODUCT RECALL URGENT - MEDICAL DEVICE CORRECTIVE ACTION" letter dated September 19, 2011, to its customers. The letter describes the product, problem and actions to be taken. Maquet service team will contact the customers within two weeks of receiving this letter to schedule an on-site service of the affected equipment. If you have any questions, call technical support at our toll free number at 1-888-627-8383.
Quantity in commerce
26 units
Distribution
Nationwide distribution.
Product codes and lots
Device Part number 66 40 440 Lot number S/N 16783-17318  510 k 041223 Device Listing No. B052264