Maquet PC Board PC1772 Spare Part Device Part number 64 67 620 Intended for treatment of and monitoring of...
Maquet Inc. began a recall of Maquet PC Board PC1772 Spare Part Device Part number 64 67 620 Intended for treatment of and monitoring of patients in range of neonates, infants and adults... on . The reason FDA records is "Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical...". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-0354-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Maquet PC Board PC1772 Spare Part Device Part number 64 67 620 Intended for treatment of and monitoring of patients in range of neonates, infants and adults with respiratory failure of respiratory insufficiency; used only by healthcare providers in hospitals or healthcare facilities and for in hospital transport.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Servo-I and Servo--S ventilator systems are equipped with a specific revision of a printed circuit board (PC1772 revision 22) that includes a component that has generated a number of nuisance alarms due to technical error message 24 (TE 24).
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- The firm, Maquet Getinge Group, sent an " PRODUCT RECALL URGENT - MEDICAL DEVICE CORRECTIVE ACTION" letter dated September 19, 2011, to its customers. The letter describes the product, problem and actions to be taken. Maquet service team will contact the customers within two weeks of receiving this letter to schedule an on-site service of the affected equipment. If you have any questions, call technical support at our toll free number at 1-888-627-8383.
- Quantity in commerce
- 10 units
- Distribution
- Nationwide distribution.
- Product codes and lots
- 510 k 041223 Device Listing No. B052264