Z-1957-2012Class II

Central Control Module for System 1: System 1 Base, 100/120V The Terumo Advanced Perfusion System 1 is...

Terumo Cardiovascular Systems Corporation / initiated 2012-06-22 / Terminated / root cause: design

Terumo Cardiovascular Systems Corporation began a recall of Central Control Module for System 1: System 1 Base, 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the... on . FDA classified it as Class II and gives the reason as "Terumo CVS has received reports of the Central Control Monitor (CCM) for Terumo Advanced Perfusion System 1 losing partial or full functionality. Symptoms that have been associated with a malfunctioning CCM include...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1957-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Central Control Module for System 1: System 1 Base, 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Terumo CVS has received reports of the Central Control Monitor (CCM) for Terumo Advanced Perfusion System 1 losing partial or full functionality. Symptoms that have been associated with a malfunctioning CCM include: full or partial loss of display, loss of control functions; inability to distinguish various status, alert, alarm conditions, the inability to operate the touch screen, or the CCM
Root cause tag
design
FDA root cause
Device Design
Action
The firm, TERUMO, sent an "URGENT MEDICAL DEVICE RECALL" notice dated June 22, 2012 and addendum to the Operator's Manual via Federal Express to its customers. The letter described the product, problem and actions to be taken. The notice states to the customers "Please note that we are not recommending that you stop using your Central Control Monitor". The customers were instructed to review the Medical Device Recall notice; assure that all users are aware of this notice; place the Addendum at the beginning of the Central Control Monitor Section (page 4-1) of the Operator's Manual and complete and return the attached Customer Response Form to the fax number indicated on the form. If you have any questions or concerns, contact Terumo CVS Customer Service at 1-800-521-2818; customer service hours are from Monday-Friday, 8 AM - 6 PM ET.
Quantity in commerce
992 units
Distribution
Worldwide distribution: USA (nationwide) and countries including: Argentina, Belgium, Canada, Chile, Colombia, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates (UAE), and Vietnam.
Product codes and lots
Catalog number: 801763 and serial numbers: 0011-0335, 0337-0610,1001-1034, 1037-1049, 1100, 1103-1106, 1109-1445, and 1447-1450.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1957-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1957-2012Class IITerminatedVoluntary: Firm initiated
Z-1958-2012Class IITerminatedVoluntary: Firm initiated