Z-0389-2012
Magnus Hybrid operating table columns 1180.01A1 and 1180.01B1
Maquet Cardiovascular Us Sales, LLC / initiated 2011-09-19 / Terminated / root cause: design
Maquet Inc. began a recall of Magnus Hybrid operating table columns 1180.01A1 and 1180.01B1 on . The reason FDA records is "A component of the Magnus hybrid OR system fails to respond to a position command". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0389-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Magnus Hybrid operating table columns 1180.01A1 and 1180.01B1
- Product code
- LFQO
- Reason for recall
- A component of the Magnus hybrid OR system fails to respond to a position command.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Maquet sent a PRODUCT RECALL URGENT MEDICAL DEVICE CORRECTIVE ACTION letter dated September 19, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. To avoid the potential of a malfunction occurring prior to their scheduled Service procedure, customers were instructed to ensure that the MAGNUS table-top is positioned at the "0" pivot position and that the MAGNUS column pivot brake is secured. If it is necessary to pivot the MAGNUS table-top prior to procedure start, ensure that the column pivot brake is completely released and then completely secured when returned to "0" pivot position. Customers were instructed to ensure that all user of the affected product and other necessary persons are informed of the product correction notice. For technical support or questions call 1-888-627-8383
- Quantity in commerce
- 45 UNITS
- Distribution
- Worldwide Distribution - USA including VA, LA, MO, NC, MD, and AZ and the countries of AE, AU, CA, CH, CN, DE, DK, ES, IE, IL, SP, PL, RU, and TH.
- Product codes and lots
- 510 k EXEMPT 1180.01A1- Lot #0002, 0004, 0005, 0007-00011, 00013- 0017, 00019-00029, 00001 1180.01B1- lOT #00001, 00002, 00005-00012, 00014, 00015, 00017, 00018