Z-0861-2012

5100-03-707 -- MOST Option STEM DISTRACTOR MATL: SS NONSTERILE QTY-1 Designed to facilitate intraoperative...

Zimmer GmbH / initiated 2011-11-15 / Terminated / root cause: labeling and UDI

Zimmer Inc. began a recall of 5100-03-707 -- MOST Option STEM DISTRACTOR MATL: SS NONSTERILE QTY-1 Designed to facilitate intraoperative decisions with a variety of options for femoral and... on . The reason FDA records is "Fractured tips: Any use of force in a certain direction to the flat side of the instrument may result in fracturing of the mating geometry, creating the potential for metal debris at the surgical site". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0861-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
5100-03-707 -- MOST Option STEM DISTRACTOR MATL: SS NONSTERILE QTY-1 Designed to facilitate intraoperative decisions with a variety of options for femoral and tibial components, implant connections and fixation of the implants.
Product code
JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Reason for recall
Fractured tips: Any use of force in a certain direction to the flat side of the instrument may result in fracturing of the mating geometry, creating the potential for metal debris at the surgical site.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
Urgent Device Correction Letters, dated October 28, 2011, which included revised Surgical Technique to be used in the Zimmer Most Option Wedge Stem Distractor were sent November 15, 2011 to all Distributors requiring Correction letters be sent to all Hospital Risk Managers and Surgeons receiving devices. Directions were provided to destroy all previous versions of the surgical technique, and review the attached revised surgical technique, with warnings of proper direction of force needed for safe separation of mated components. A FAX -Back Certificate of Acknowledgement is to be sent to (574) 372-4265. Call Zimmer at 1-877-946-2761 with questions or concerns.
Quantity in commerce
70
Distribution
Worldwide Distribution -- US, including states of TX, IN, GA, WI, MN, NJ, MO, IL, VA, MI, PA, NC, OR, FL, KS, NY, AZ, CO, and UT and country of Switzerland.
Product codes and lots
Lots 1519123, 1616672, 1627174, 60281712, 60773654, 60795066, 60841064, 61061148, 61348637, and 61770544.