5100-03-707 -- MOST Option STEM DISTRACTOR MATL: SS NONSTERILE QTY-1 Designed to facilitate intraoperative...
Zimmer Inc. began a recall of 5100-03-707 -- MOST Option STEM DISTRACTOR MATL: SS NONSTERILE QTY-1 Designed to facilitate intraoperative decisions with a variety of options for femoral and... on . The reason FDA records is "Fractured tips: Any use of force in a certain direction to the flat side of the instrument may result in fracturing of the mating geometry, creating the potential for metal debris at the surgical site". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0861-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- 5100-03-707 -- MOST Option STEM DISTRACTOR MATL: SS NONSTERILE QTY-1 Designed to facilitate intraoperative decisions with a variety of options for femoral and tibial components, implant connections and fixation of the implants.
- Product code
- JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
- Reason for recall
- Fractured tips: Any use of force in a certain direction to the flat side of the instrument may result in fracturing of the mating geometry, creating the potential for metal debris at the surgical site.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- Urgent Device Correction Letters, dated October 28, 2011, which included revised Surgical Technique to be used in the Zimmer Most Option Wedge Stem Distractor were sent November 15, 2011 to all Distributors requiring Correction letters be sent to all Hospital Risk Managers and Surgeons receiving devices. Directions were provided to destroy all previous versions of the surgical technique, and review the attached revised surgical technique, with warnings of proper direction of force needed for safe separation of mated components. A FAX -Back Certificate of Acknowledgement is to be sent to (574) 372-4265. Call Zimmer at 1-877-946-2761 with questions or concerns.
- Quantity in commerce
- 70
- Distribution
- Worldwide Distribution -- US, including states of TX, IN, GA, WI, MN, NJ, MO, IL, VA, MI, PA, NC, OR, FL, KS, NY, AZ, CO, and UT and country of Switzerland.
- Product codes and lots
- Lots 1519123, 1616672, 1627174, 60281712, 60773654, 60795066, 60841064, 61061148, 61348637, and 61770544.