The EnCor biopsy probe, designed for use with EnCor and EnCor Enspire Vacuum Assisted Biopsy (VAB) driver...
Bard Peripheral Vascular Inc began a recall of The EnCor biopsy probe, designed for use with EnCor and EnCor Enspire Vacuum Assisted Biopsy (VAB) driver systems. The EnCor Biopsy Device is indicated for use... on . The reason FDA records is "This recall was voluntarily initiated by the firm due to the product 's damaged packaging trays; cracked EnCor (Trademark) packaging trays, plastic PETG tray the probe is packaged in was cracked". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0351-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- The EnCor biopsy probe, designed for use with EnCor and EnCor Enspire Vacuum Assisted Biopsy (VAB) driver systems. The EnCor Biopsy Device is indicated for use in acquiring tissue for diagnosing breast abnormalities and is supplied sterile and intended for single use.
- Product code
- FCK Instrument, Biopsy, Suction
- Reason for recall
- This recall was voluntarily initiated by the firm due to the product 's damaged packaging trays; cracked EnCor (Trademark) packaging trays, plastic PETG tray the probe is packaged in was cracked.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- The firm sent the customer notification letter on 11/07/11, via FedEx with proof of delivery notice. The letter notified customers of the reason for recall that units from the affected lot may have packaging damage resulting from transportation. The letter instructed customers to check all inventory locations , remove any identified product, to not use the affected product, and return the affected product to Bard Peripheral Vascular, Inc., 1415 W. 3rd Street, Tempe, Arizona 85281. The letter also instructed customers to fill out the recall and effectiveness check form and fax it to 1-800-894-6772. Customers were instructed to call 1-800-321-4254, Option #2 Ext 2727 for a return authorization number to facilitate the expedient return of the product. For questions regarding this recall call 480-894-9515
- Quantity in commerce
- 64570 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Canada, Europe, LAPAC, and Australia
- Product codes and lots
- ECP0110G, ECP017G, ECP0112G, ECPMR0110G, ECP017GV, ECPMR0110GBT, ECP0112GV, ECPMR017G.