Step 2, 3000-22000 GPD (Sodium Hydroxide 2.0N) cleaning solution. 4 oz plastic bottle. Lot # 11IC002, Serial...
Isopure Corp began a recall of Step 2, 3000-22000 GPD (Sodium Hydroxide 2.0N) cleaning solution. 4 oz plastic bottle. Lot # 11IC002, Serial #'s: 11IC00210 - 11IC002294. The Step 2, 3000-2000... on . The reason FDA records is "The cleaning solution Step 2, 3000-22000 GPD labeled bottles (NaOH) were filled with Step 1, 3000-22000 GPD chemical (HCl)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0190-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Isopure Corp
- Product
- Step 2, 3000-22000 GPD (Sodium Hydroxide 2.0N) cleaning solution. 4 oz plastic bottle. Lot # 11IC002, Serial #'s: 11IC00210 - 11IC002294. The Step 2, 3000-2000 GPD chemical is used in the cleaning procedures for the Isopure MD400 Series Reverse Osmosis System. Step
- Product code
- KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
- Reason for recall
- The cleaning solution Step 2, 3000-22000 GPD labeled bottles (NaOH) were filled with Step 1, 3000-22000 GPD chemical (HCl).
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Isopure sent a Product Recall Notice dated October 24, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue the use of any Step 2 chemical from the affected product, place any remaining chemical into the original box, seal it with tape, and segregate the product to prevent it from being placed back into stock. The firm will pick up the product for return on a call tag. Customers would receive replacement product for all affected cases of the Step 2 chemical the firm received. For questions call 502-722-1000.
- Quantity in commerce
- 91cs/12/4 oz plastic bottles
- Distribution
- Nationwide Distribution including the following states: AL, AZ, CA, FL, GA, IN, KS, LA, MA, MO, MS, NC, NH, NJ, NM, NV, OH, OR, PA, SC, TN, TX, and VA
- Product codes and lots
- Lot # 11IC002, Serial #'s: 11IC00210 - 11IC002294.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.