Coulter LH500 Series Analyzer, Part number: 178832, 178833, 178834. LH 500 Series System Instructions for...
Beckman Coulter Inc. began a recall of Coulter LH500 Series Analyzer, Part number: 178832, 178833, 178834. LH 500 Series System Instructions for Use, PN 624602 Rev BA. The COULTER LH 500 Analyzer is... on . The reason FDA records is "The recall was initiated because Beckman has identified a problem with Integrated Circuit chips used in different boards throughout the analyzer". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0599-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Coulter LH500 Series Analyzer, Part number: 178832, 178833, 178834. LH 500 Series System Instructions for Use, PN 624602 Rev BA. The COULTER LH 500 Analyzer is a quantitative, automated hematology analyzer and leukocyte differential cell counter For In Vitro Diagnostic Use in clinical laboratories. The LH500 Analyzer also provides a semi-automated reticulocyte analysis.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- The recall was initiated because Beckman has identified a problem with Integrated Circuit chips used in different boards throughout the analyzer.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Beckman Coulter sent an "URGENT PRODUCT CORRECTION" letter dated October 13, 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers were instructed to contact Beckman Coulter Service at (800) 526-7694 if they were experiencing failures. Customers not experiencing failures, but their system may have the suspect chips, were told they would be contacted by Beckman Coulter Service to correct their problem. Customers were also instructed to complete and return an enclosed Response Form.
- Quantity in commerce
- 117 units total
- Distribution
- Worldwide Distribution.
- Product codes and lots
- N/A