Z-0463-2012
Smith & Nephew 5.5mm DYONICS BONECUTTER ELECTROBLADE Resector Catalog Number: 7205962
Smith & Nephew, Inc. / initiated 2011-11-07 / Terminated / root cause: manufacturing process
Smith & Nephew, Inc. Endoscopy Division began a recall of Smith & Nephew 5.5mm DYONICS BONECUTTER ELECTROBLADE Resector Catalog Number: 7205962 on . The reason FDA records is "Manufacturing issue: The distal tip of the outer sheath (return electrode) may detach and become loose in the joint during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0463-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Smith & Nephew 5.5mm DYONICS BONECUTTER ELECTROBLADE Resector Catalog Number: 7205962.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Manufacturing issue: The distal tip of the outer sheath (return electrode) may detach and become loose in the joint during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Smith & Nephew Inc. Endoscopy Division notified facilities on 11/7/11 via an "Urgent - Product Recall 1st Notification - Urgent" letter and/or phone. The letter identified the affected product and the reason for the recall. It also discussed the potential risks, customer actions, and instructions for the return of the product. Customers were to check their inventory and locate any of the affected product. They were to also complete the requested contact information and return the notification letter. In addition, customers were to contact Smith & Nephew Endoscopy Division Returns Group at [email protected] for information on how to return the affected product and receive replacement product. Questions, contact Hoangthi Le directly by phone at 508-337-3731.
- Quantity in commerce
- 1743 units
- Distribution
- Worldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Dubai, Finland, France, Germany, Norway, South Africa, Switzerland, United Kingdom, and Venezuela.
- Product codes and lots
- Lot Numbers: 830567, 830568, 830569, 863004, 868509, 868519, 875226, 875227, and 897191.