Smith & Nephew 5.5mm DYONICS BONECUTTER ELECTROBLADE Resector Catalog Number: 7205962

Smith & Nephew, Inc. / initiated 2011-11-07 / Terminated / root cause: manufacturing process

Smith & Nephew, Inc. Endoscopy Division began a recall of Smith & Nephew 5.5mm DYONICS BONECUTTER ELECTROBLADE Resector Catalog Number: 7205962 on . The reason FDA records is "Manufacturing issue: The distal tip of the outer sheath (return electrode) may detach and become loose in the joint during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0463-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Smith & Nephew 5.5mm DYONICS BONECUTTER ELECTROBLADE Resector Catalog Number: 7205962.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Manufacturing issue: The distal tip of the outer sheath (return electrode) may detach and become loose in the joint during use.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Smith & Nephew Inc. Endoscopy Division notified facilities on 11/7/11 via an "Urgent - Product Recall 1st Notification - Urgent" letter and/or phone. The letter identified the affected product and the reason for the recall. It also discussed the potential risks, customer actions, and instructions for the return of the product. Customers were to check their inventory and locate any of the affected product. They were to also complete the requested contact information and return the notification letter. In addition, customers were to contact Smith & Nephew Endoscopy Division Returns Group at [email protected] for information on how to return the affected product and receive replacement product. Questions, contact Hoangthi Le directly by phone at 508-337-3731.
Quantity in commerce
1743 units
Distribution
Worldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Dubai, Finland, France, Germany, Norway, South Africa, Switzerland, United Kingdom, and Venezuela.
Product codes and lots
Lot Numbers: 830567, 830568,  830569,  863004,  868509,  868519,  875226,  875227, and 897191.