Z-0358-2012
Integra Surgical Luxtec MLX Light Source Catalog Numbers: 00MLXEU
Integra LifeSciences Corporation / initiated 2011-10-14 / Terminated / root cause: manufacturing process
Integra LifeSciences Corp. began a recall of Integra Surgical Luxtec MLX Light Source Catalog Numbers: 00MLXEU on . The reason FDA records is "A loose screw may result in shifting of components and improper function of the turret or actuator assembly". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0358-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Integra LifeSciences Corporation
- Product
- Integra Surgical Luxtec MLX Light Source Catalog Numbers: 00MLXEU
- Product code
- FCW Light Source, Fiberoptic, Routine
- Reason for recall
- A loose screw may result in shifting of components and improper function of the turret or actuator assembly.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Integra sent an URGENT: Field Corrective Action letter dated October 14, 2011 to all affected customers via FedEx. The letter identified the affected product, the problem, and the actions to be taken with instructions to immediately inspect their inventory for potentially loose screws within the affected unit(s). If the unit(s) were observed with loose screws, the unit(s) were to be returned to Integra Surgical for repair and servicing. Customers were instructed to complete the Acknowledgement and Return Form per the instructions on the form. For questions call Integra Surgical Customer Service at 1-800-431-1123.
- Quantity in commerce
- 19
- Distribution
- Worldwide Distribution - USA (nationwide) including the states of: Arizona, California, Colorado, Florida, Indiana, Mississippi, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Montana, Washington, Wyoming, and Puerto Rico, and the countries of: Australia, Germany, Italy, Japan, Netherlands, New Zealand, Pakistan, Sweden, Switzerland, and Turkey.
- Product codes and lots
- SN'S: P0102011 P0203911 P0204511 P0302011 P03011211 P0303411 P0104911 P0105211 P0105811 P0106611