SUPER TORQUE MB Angiographic Catheter product is labeled in part: "***5F (1.65 mm)***110 cm .038" (0.97...
Cordis Corporation began a recall of SUPER TORQUE MB Angiographic Catheter product is labeled in part: "***5F (1.65 mm)***110 cm .038" (0.97 mm)***PIG***Max. 1200 psi (8274 kPa)***Markers***Open... on . The reason FDA records is "Cordis Corporation is issuing a Correction for their SUPER TORQUE MB Angiographic Catheter after receiving a report of marker bands dislodged during a procedure. Model Numbers 532598A, 532598B, 532598C, 532598D...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0204-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cordis US Corp.
- Product
- SUPER TORQUE MB Angiographic Catheter product is labeled in part: "***5F (1.65 mm)***110 cm .038" (0.97 mm)***PIG***Max. 1200 psi (8274 kPa)***Markers***Open End 6 Sideholes***Angiographic Catheter*** REF Cat No.***Lot***Use By***Assembled in Mexico Cordis Corporation, 14201 NW 60th Ave. Miami Lakes, Florida 33014, USA***EC***REP*** Cordis Cashel, Cahir Road Cashel, Co Tipperary, Ireland***Cordis***a Johnson&Johnson company***" UPN H739532598A3, Cat No 532-598A, No. MB 10, French Size 5, Shape (PIG), Length cm 110, Flow Rate, ml/sec, 18, Sideholes 6. UPN H739532598B3, Cat No 532-598B, No. MB 20, French Size 5, Shape (PIG), Length cm 110, Flow Rate, ml/sec, 18, Sideholes 6. UPN H73953298C3, Cat No 532-598C, No MB 20 French Size 5, Shape (PIG), Length, cm 65, Flow Rate, ml/sec, 30, Side Holes 6. UPN H73953298D3, Cat No 532-598D, No MB 2 French Size 5, Shape (PIG), Length , cm 70, Flow Rate, ml/sec, 30, Side Holes 10. Cordis SUPER TORQUE MB angiographic catheters with Market Bands are designed to provide angiographic visualization and linear measurement of the vasculature when combined with the delivery of radiopaque contrast media to selected sites in the vascular system.
- Product code
- DQO Catheter, Intravascular, Diagnostic
- Reason for recall
- Cordis Corporation is issuing a Correction for their SUPER TORQUE MB Angiographic Catheter after receiving a report of marker bands dislodged during a procedure. Model Numbers 532598A, 532598B, 532598C, 532598D, SRD5724MB and SRD5727MB are affected by the Cordis Correction.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Cordis Corporation sent a Correction letter dated November 21, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers are instructed to read the "Description" and "Recommendations" sections, sign and return an enclosed Acknowledgement Form, and to maintain awareness of the communication until the information has been incorporated into the Cordis Super Torque MB Catheter labeling. Contact Cordis at 800-781-0282 for questions regarding this notice.
- Quantity in commerce
- 118,690 units
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico.
- Product codes and lots
- Not Lot Specific. Catalog No. 532-598A, 532-598B, 532-598C, and 532-598D.