Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0205-2012

SUPER TORQUE MB Angiographic Catheter product is labeled in part: "***5F (1.65 mm)***110 cm .038" (0.97...

Cordis US Corp. / initiated 2011-11-21 / Terminated / root cause: labeling and UDI

Cordis Corporation began a recall of SUPER TORQUE MB Angiographic Catheter product is labeled in part: "***5F (1.65 mm)***110 cm .038" (0.97 mm)***MULTIPURPOSE A SUPER TORQUE MB***Max. 1200 psi... on . The reason FDA records is "Cordis Corporation is issuing a Correction for their SUPER TORQUE MB Angiographic Catheter after receiving a report of marker bands dislodged during a procedure. Model Numbers 532598A, 532598B, 532598C, 532598D...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0205-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cordis US Corp.
Product
SUPER TORQUE MB Angiographic Catheter product is labeled in part: "***5F (1.65 mm)***110 cm .038" (0.97 mm)***MULTIPURPOSE A SUPER TORQUE MB***Max. 1200 psi (8274 kPa)***Markers***OPEN end***Sideholes***SPECIAL***Angiographic Catheter*** REF Cat No.***Lot***Use By***Sterile EO***Assembled in Mexico Cordis Corporation, 14201 NW 60th Ave. Miami Lakes, Florida 33014, USA***EC***REP*** Cordis Cashel, Cahir Road Cashel, Co Tipperary, Ireland***Cordis***a Johnson&Johnson company***" UPN SRD5724MB, No MB 2, Length, cm 100, Sideholes 2. UPN SRD5727MB, No MB 4, Length, cm 80, Sideholes 4. Cordis SUPER TORQUE MB angiographic catheters with Market Bands are designed to provide angiographic visualization and linear measurement of the vasculature when combined with the delivery of radiopaque contrast media to selected sites in the vascular system.
Product code
DQO Catheter, Intravascular, Diagnostic
Reason for recall
Cordis Corporation is issuing a Correction for their SUPER TORQUE MB Angiographic Catheter after receiving a report of marker bands dislodged during a procedure. Model Numbers 532598A, 532598B, 532598C, 532598D, SRD5724MB and SRD5727MB are affected by the Cordis Correction.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Cordis Corporation sent a Correction letter dated November 21, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers are instructed to read the "Description" and "Recommendations" sections, sign and return an enclosed Acknowledgement Form, and to maintain awareness of the communication until the information has been incorporated into the Cordis Super Torque MB Catheter labeling. Contact Cordis at 800-781-0282 for questions regarding this notice.
Quantity in commerce
310 units
Distribution
Worldwide Distribution-USA (nationwide) including Puerto Rico.
Product codes and lots
Not Lot Specific. Catalog No. SRD5724MB and SRD5727MB.