Z-0866-2012

Gas cylinders used in Gunnell manual wheelchairs with dynamic recline feature, including: Bariatric Rehab...

GUNNELL, Inc. / initiated 2011-11-01 / Terminated / root cause: design

Gunnell, Inc. began a recall of Gas cylinders used in Gunnell manual wheelchairs with dynamic recline feature, including: Bariatric Rehab Tough & Tilt (BTNT) Product Usage: The purpose of... on . The reason FDA records is "Gunnell has received complaints that the gas cylinders used in Gunnell manual wheelchairs would not hold a firm position, or as described by users, the back support or canes were spongy in the Rehab Tough and Tilt...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0866-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GUNNELL, Inc.
Product
Gas cylinders used in Gunnell manual wheelchairs with dynamic recline feature, including: Bariatric Rehab Tough & Tilt (BTNT) Product Usage: The purpose of these wheelchairs is to provide mobility to individuals limited to a seating position. They may also be used as attendant propelled transport devices in a health care environment such as a hospital, nursing home or extended care facility.
Product code
IOR Wheelchair, Mechanical
Reason for recall
Gunnell has received complaints that the gas cylinders used in Gunnell manual wheelchairs would not hold a firm position, or as described by users, the back support or canes were spongy in the Rehab Tough and Tilt (TNT), Bariatric Rehab Tough and Tilt (BTNT), Rehab Recline and Mobility (RAM), Bariatric Rehab Recline and Mobility (BRAM), and Rehab Kidster.
Root cause tag
design
FDA root cause
Component design/selection
Action
Gunnell, Inc. sent a Medical Device Correction letters dated November 1, 2011 to all consignees. The letter identified the product, problem, and actions to be taken. The letter states that Gunnell Inc will provide free replacement cylinders and reimburse for each correction verified. The notice instructs customers to complete and return the Response Form. Customers were instructed to notify their customers by including a copy of the correction notice. Question and concerns should be addressed to [email protected]. or call 1-800-551-0055, attention Andy Evancho.
Quantity in commerce
10 units
Distribution
Worldwide Distribution - USA (nationwide) nad the countries of: Australia and Canada
Product codes and lots
Serial numbers: 32031-0504, 32280-0704, 34402-0207, 33383-01 06, 33304-1205, 33409-0106, 33436-0206, 33515-0306, 34712-0607, 35899-0209