TrueBeam and TrueBeam STx, Model Number: H19, Ref/FSCA identifier: CP-06381 are intended to provide...
Varian Medical Systems, Inc. Oncology Systems began a recall of TrueBeam and TrueBeam STx, Model Number: H19, Ref/FSCA identifier: CP-06381 are intended to provide stereotactic radiosurgery and precision radiotherapy for... on . The reason FDA records is "An anomaly has been identified with the TrueBeam and TrueBeam STx systems where, under certain tuning conditions, the electron beam emerging from the bend magnet may have an elongated spot shape". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0364-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- TrueBeam and TrueBeam STx, Model Number: H19, Ref/FSCA identifier: CP-06381 are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
- Product code
- -IYE
- Reason for recall
- An anomaly has been identified with the TrueBeam and TrueBeam STx systems where, under certain tuning conditions, the electron beam emerging from the bend magnet may have an elongated spot shape.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 11/04/2011, an Urgent Medical Device Correction Letter was distributed to all affected users, with a description of the problem and user corrective action steps. Varian is developing a corrective action for this issue. A customer service representative will contact you when the correction is available to schedule installation on your system. Questions regarding the letter should be directed to the Varian Oncology Help Desk Contact Information at USA and Canada: 1-888-827- 4265 or Europe +41-41-749-8844.
- Distribution
- Worldwide distribution including the US and Internationally.
- Product codes and lots
- Model numbers: H191113, H191034, H191105, H191154, H191046, H191053, H191044, H191015, H191035, H191108, 191152, H191140, H191069, H191017, H191027, H191109, 191130, H191002, H191071, H191024, H191078, H191111, 191072, H191102, H191074, H191087, H191151, H191125, 191018, H191058, H191038, H191115, H191085, H191032, 191077, H191023, H191103, H191095, H191163, H191146, 191124, H191138, H191093, H191097, H191040, H191030, 191029, H191060, H191066, H191122, H191054, H191126, 191114, H191082, H191067, H191081, H191005, H191013, 191116, H191051, H191094, H191052, H191031, H191021, 191048, H191139, H191099, H191020, H191039, H191128, 191156, H191057, H191061, H191112, H191049, H191083, 191008, H191123, H191065, H191119, H191089, H191120, 191092, H191134, H191129, H191056, H191090, H191033, 191047, H191037, H191076, H191007, H191028, H191080, 191106, H191118, H191073, H191117, H191155, H191086, 191141, H191025, H191036, H191042, H191079, H191135, 191010, H191100, H191107, H191127, H191063, H191070, 191153, H191026, H191096, H191014, H191064, H191059, 191009, H191142, H191068, H191001, H191132, H191011, 191003, H191136, H191022, H191121, H191016, H191012, 191045, H191101, H191088, H191041, H191043, H191150, 191019, H191055, H191133, H191131, H191091, H191098, 191006 and H191110