SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. SoftReports is a web-based report...
SCC Soft Computer began a recall of SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. SoftReports is a web-based report launching tool to be used in a hospital or laboratory... on . The reason FDA records is "On 10/25/2011 SCC Soft Computer, Clearwater, FL initiated a correction on the SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. When SoftReports Designer opened a layout, the system did not load the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0362-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SCC Soft Computer
- Product
- SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. SoftReports is a web-based report launching tool to be used in a hospital or laboratory environment by knowledgeable, trained, and experienced personnel. SoftReports enables the user to schedule and generate reports for the purpose of viewing and printing data stored in SCC product databases.
- Product code
- LNX
- Reason for recall
- On 10/25/2011 SCC Soft Computer, Clearwater, FL initiated a correction on the SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. When SoftReports Designer opened a layout, the system did not load the layout with the previously saved value in the NewPage property. This caused multiple patient records to be printed on the same page. All affected clients were notified on the issue on
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- SCC Soft Computer sent a "Correction Notification" letter dated October 25, 2011 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. The hot fixes that were previously loaded which introduced this issue were corrected and delivered to all affected customers as a mandatory correction in hot fixes 1.7199.8 and 1.7200.8 and higher.
- Quantity in commerce
- 5
- Distribution
- Nationwide Distribution.
- Product codes and lots
- SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7.