Z-0523-2012
MAMMOMAT Inspiration Full field digital mammographic x-ray system
Siemens Medical Solutions USA, Inc. / initiated 2011-09-15 / Terminated
Siemens Medical Solutions USA, Inc began a recall of MAMMOMAT Inspiration Full field digital mammographic x-ray system on . The reason FDA records is "Lowering the swivel arm with "object table" while an operator's or patient's body parts are underneath the "object table" may cause the "object table" to collide with and injure the operator or patient. Therefore...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0523-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- MAMMOMAT Inspiration Full field digital mammographic x-ray system
- Product code
- MUE Full Field Digital, System, X-Ray, Mammographic
- Reason for recall
- Lowering the swivel arm with "object table" while an operator's or patient's body parts are underneath the "object table" may cause the "object table" to collide with and injure the operator or patient. Therefore attention should be paid when adjusting "object table" height to avoid a possible risk of injury to the operator or the patient. The movement of the swivel arm may also cause an injur
- FDA root cause
- Use error
- Action
- Siemens Medical Solutions USA, Inc sent a Safety Advisory Notice dated January 8, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. The notification included an Addendum to the User Manual. We appreciate your understanding and cooperation with this safety advisory and ask you to immediately instruct your operating personnel accordingly. Please ensure that this safety advisory is placed in the System's Operator Manual. Your operating personnel should maintain awareness over an appropriate defined period. Further questions please call (610) 219-4834.
- Quantity in commerce
- 12
- Distribution
- Nationwide Distribution including the states of FL, MD, MI, NE, NY, NC, ND, OH, and SC.
- Product codes and lots
- Model number 10140000