Z-0522-2012

syngo.plaza Image processing system

Siemens Medical Solutions USA, Inc. / initiated 2011-09-21 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of syngo.plaza Image processing system on . The reason FDA records is "Potential malfunction when using syngo.plaza. On syngo.plaza workplaces large studies can be loaded into the viewer if transferred from a DICOM node or imported even while image transfer is not yet completed. In these...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0522-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
syngo.plaza Image processing system
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Potential malfunction when using syngo.plaza. On syngo.plaza workplaces large studies can be loaded into the viewer if transferred from a DICOM node or imported even while image transfer is not yet completed. In these situations more images of that study may exist which are not shown in the syngo.plaza viewer.
Root cause tag
software
FDA root cause
Software design
Action
Siemens Medical Solutions USA sent a Customer Safety Advisory Notice dated March 10, 2011 to the affected customer. The letter identified the product the problem and the action needed to be taken by the customer. The letter informed the customer of the potential issues and provides additional information on the implemented resolution. For further questions please call (610) 219-4834.
Quantity in commerce
1
Distribution
Nationwide Distribution New York
Product codes and lots
Model number 10592457