Z-0522-2012
syngo.plaza Image processing system
Siemens Medical Solutions USA, Inc. / initiated 2011-09-21 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of syngo.plaza Image processing system on . The reason FDA records is "Potential malfunction when using syngo.plaza. On syngo.plaza workplaces large studies can be loaded into the viewer if transferred from a DICOM node or imported even while image transfer is not yet completed. In these...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0522-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- syngo.plaza Image processing system
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Potential malfunction when using syngo.plaza. On syngo.plaza workplaces large studies can be loaded into the viewer if transferred from a DICOM node or imported even while image transfer is not yet completed. In these situations more images of that study may exist which are not shown in the syngo.plaza viewer.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens Medical Solutions USA sent a Customer Safety Advisory Notice dated March 10, 2011 to the affected customer. The letter identified the product the problem and the action needed to be taken by the customer. The letter informed the customer of the potential issues and provides additional information on the implemented resolution. For further questions please call (610) 219-4834.
- Quantity in commerce
- 1
- Distribution
- Nationwide Distribution New York
- Product codes and lots
- Model number 10592457