Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0367-2012

Zimmer ZMR Femoral Spout & Cone Bodies and Trabecular Metal Hip Prosthesis Femoral Stems: FEMORAL BODY...

Zimmer GmbH / initiated 2011-11-16 / Terminated / root cause: manufacturing process

Zimmer Inc. began a recall of Zimmer ZMR Femoral Spout & Cone Bodies and Trabecular Metal Hip Prosthesis Femoral Stems: FEMORAL BODY REVISION NITRIDED POROUS 12/14 NECK TAPER EXTENDED 46 MM... on . The reason FDA records is "Lack of assurance of proper metal fatigue strength due to metal grain structure anomaly. In 2006 a recall was initiated as a result of the grain anomaly; however, these 4 lots had been split off during normal production...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0367-2012
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer ZMR Femoral Spout & Cone Bodies and Trabecular Metal Hip Prosthesis Femoral Stems: FEMORAL BODY REVISION NITRIDED POROUS 12/14 NECK TAPER EXTENDED 46 MM NECK OFFSET SPOUT BODY SIZE E 55 MM BUILD-UP TIVANIUM TI-6AL-4V ALLOY STERILE. Item number: 00-9991-021-55
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
Lack of assurance of proper metal fatigue strength due to metal grain structure anomaly. In 2006 a recall was initiated as a result of the grain anomaly; however, these 4 lots had been split off during normal production and had not been included in the 2006 recall.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 11/16/11, Zimmer sent letters titled "URGENT MEDICAL DEVICE RECALL" to distributors via electronic mail and to risk managers and surgeons via FedEx indicating that the product must be immediately removed from use and quarantined. " Stop using the affected lots and place affected product in quarantine. " Review available product on the shelf " Identify those part/lot combinations listed in the recall communication " Immediately remove any implants found that match part/lot combinations listed in the recall communication and quarantine " Return affected lots and signature of acknowledgement to Zimmer
Quantity in commerce
1 unit (International Only Switzerland)
Distribution
Worldwide distribution: USA (NY), AUSTRALIA, SWITZERLAND, and DENMARK
Product codes and lots
Item number: 00-9991-021-55 Lot number: 00104623