SoftLab GUI versions 4.0.2.8-4.0.2.10; 4.0.3.12-4.0.3.13; and 4.0.4.1-4.0.4.4. SoftLab is a laboratory...
SCC Soft Computer began a recall of SoftLab GUI versions 4.0.2.8-4.0.2.10; 4.0.3.12-4.0.3.13; and 4.0.4.1-4.0.4.4. SoftLab is a laboratory information system to be used in a medical research or... on . The reason FDA records is "On 05/03/2011 SCC Soft Computer, Clearwater, FL initiated a correction on the following SoftLab GUI versions 4.0.2.8-4.0.2.10; 4.0.3.12-4.0.3.13; and 4.0.4.1-4.0.4.4. A client reported the PT and INR results did not get...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0277-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SCC Soft Computer
- Product
- SoftLab GUI versions 4.0.2.8-4.0.2.10; 4.0.3.12-4.0.3.13; and 4.0.4.1-4.0.4.4. SoftLab is a laboratory information system to be used in a medical research or clinical laboratory by knowledgeable, trained, and experienced personnel. SoftLab enables the user to handle data-processing and laboratory activities, for the purpose of storing, managing, querying, and report laboratory data.
- Product code
- LNX
- Reason for recall
- On 05/03/2011 SCC Soft Computer, Clearwater, FL initiated a correction on the following SoftLab GUI versions 4.0.2.8-4.0.2.10; 4.0.3.12-4.0.3.13; and 4.0.4.1-4.0.4.4. A client reported the PT and INR results did not get flagged. The PROT should have been flagged as HIGH and the INR should have been flagged as PANIC. Neither test was flagged. All affected clients were notified of the issue on 05/03
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- SCC Soft Computer initiated a Field correction on May 3, 2011. All customers with affected software have been notified of this issue. Clients were instructed to have a "valcheck" utility run to determine if there are any tests in their system affected by the issue. They were supposed to indicate whether they wanted to schedule to receive the patch or hot fix. If you have any further questions please call ( 727 ) 789-0100.
- Quantity in commerce
- 69
- Distribution
- Worldwide Distribution-- USA ( nationwide) and Canada.
- Product codes and lots
- SoftLab GUI versions 4.0.2.8-4.0.2.10; 4.0.3.12-4.0.3.13; and 4.0.4.1-4.0.4.4.