CyberKnife Robotic Radiosurgery System: Lung Optimized Treatment option. The CyberKnife treatment is...
Accuray Incorporated began a recall of CyberKnife Robotic Radiosurgery System: Lung Optimized Treatment option. The CyberKnife treatment is indicated for the treatment planning and image guided... on . The reason FDA records is "User Facility reported an anomaly where inhale and exhale CT pairs used for treatment planning did not represent the same magnitude of respiration that was displayed during treatment delivery". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0382-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Accuray Incorporated
- Product
- CyberKnife Robotic Radiosurgery System: Lung Optimized Treatment option. The CyberKnife treatment is indicated for the treatment planning and image guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- User Facility reported an anomaly where inhale and exhale CT pairs used for treatment planning did not represent the same magnitude of respiration that was displayed during treatment delivery.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Accuray sent a Urgent Medical Device Correction letter dated November 7, 2011, via Federal Express to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The information provided in this letter will be incorporated in user manuals shipped with the CyberKnife system. If you have any questions or concerns regarding this issue, please do not hesitate to contact Accuray Customer Support at 1. (877) 668-8667 ( USA) 1. (408) 716-4700 ( non USA )
- Quantity in commerce
- 30
- Distribution
- Worldwide Distribution -- USA ( nationwide ) including the states of FL, TN, IL, NJ, CO, PA, MD, CA, MS, MN, SC, WA, OK, GA, AZ, and TX., and the countries of France and United Kingdom.
- Product codes and lots
- Model number: PN 032384