Z-0396-2012
OR table TITAN T, Surgical Table. Surgical bed with Orthopedic Accessories
Hill-Rom Inc. / initiated 2011-01-21 / Terminated / root cause: design
Trumpf Medical Systems, Inc. began a recall of OR table TITAN T, Surgical Table. Surgical bed with Orthopedic Accessories on . The reason FDA records is "Product is being recalled due to unintended movement on the Titan Tables". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0396-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- OR table TITAN T, Surgical Table. Surgical bed with Orthopedic Accessories.
- Product code
- FQO Table, Operating-Room, Ac-Powered
- Reason for recall
- Product is being recalled due to unintended movement on the Titan Tables.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Trumpf Medical Systems sent an initial notification to customers dated January 24, 2011. A subsequent "URGENT SAFETY NOTICE-UPDATE" letter dated May 26, 2011 was sent to all affected customers. The letter described the product, problem, and actions to be taken by the customers. A Corrective Action Plan includes a software update to be installed by a Trumpf Technical Service Representative. Contact Technical Service at 888-474-9359 for questions regarding this notice.
- Quantity in commerce
- see line 1.
- Distribution
- Worldwide Distribution.
- Product codes and lots
- Material number: 197903, All Serial numbers.