Varian brand Eclipse Release 10 with RT Chart Release 10 (Client Build 10.0.28 or Build 10.0.34), An Eclipse...
Varian Medical Systems, Inc. Oncology Systems began a recall of Varian brand Eclipse Release 10 with RT Chart Release 10 (Client Build 10.0.28 or Build 10.0.34), An Eclipse Treatment Planning System used with ARIA Radiation... on . The reason FDA records is "An anomaly has been identified with Varian Eclipse and RT Chart where adding a reference point in RT Chart and editing the field dose contributions may change the Prescribed Dose Per Fraction in Eclipse". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0370-2012
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Varian brand Eclipse Release 10 with RT Chart Release 10 (Client Build 10.0.28 or Build 10.0.34), An Eclipse Treatment Planning System used with ARIA Radiation Oncology (RT Chart module), Reference/FSCA Identifier: CP-05277, Model Number: H48; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- An anomaly has been identified with Varian Eclipse and RT Chart where adding a reference point in RT Chart and editing the field dose contributions may change the Prescribed Dose Per Fraction in Eclipse.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Varian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION/URGENT FIELD SAFETY NOTICE" dated October 31, 2011 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. Varian has corrected the issue and will contact customers to schedule installation of the corrected version of the system. Contact the Varian Oncology Help Desk at 1-888-827-4265 for questions regarding this notice. .
- Distribution
- Worldwide Distribution.
- Product codes and lots
- CODES: All codes from H480005 through H48C896